Description
Overview
The EMC Pro Plus made by Thermo Fisher Scientific is a Conducted Immunity Generator designed for 6kV surge/combination wave, 4.4kV EFT/Burst, 100kHz ringwave at 6kV, and dips testing.This test system provides a single unit solution for surge, EFT, ringwave, and dips and interruptions and is upgraded to meet the latest versions of these standards. This test equipment comes with an A2LA ISO 17025 accredited calibration and free technical support. We also offer rentals of the 10x700 Telecom wave board providing capabilities with the EMC Pro Plus platform.
This EMI Immunity Generator is an easy to use, cost effective solution to a diverse set of testing needs. This system provides easy to use front panel display and can be controlled via computer through windows based software. It also monitors output of the coupling unit and automatically switches connections according to coupling mode.
The EMC Pro Plus is designed to test for a variety of EMC immunity industry standards including EN/IEC 61000-4-4, EN/IEC 61000-4-5, EN/IEC 61000-4-11, EN/IEC 61000-4-12, CEMark, and ANSI IEEE C62.41 cat. B. The EMC Pro Plus rental equipment It also can also be configured for custom test protocols to meet specific test requirements using the software and a full line of options and accessories.
Data SheetUser ManualSpecifications
Common Standards |
EN/IEC 61000-4-4, EN/IEC 61000-4-5, EN/IEC 61000-4-11, EN/IEC 61000-4-12, CEMark, ANSI IEEE C62.41 cat. B |
Integrated CDN: |
|
AC Voltage and Current |
50-250 VAC, 16A max AC @250V |
DC Voltage and Current |
100 VDC max, 10A max to 100V, 16A max to 48V |
EFT Burst 5/50ns: |
|
Standards |
IEC 61000-4-4, ed.3 |
Voltage Waveform into 50Ω load |
5ns Rise Time +/-1.5ns, 50µs pulse width +/-15ns |
Voltage Waveform into 1000Ω load |
5ns Rise Time +/-1.5ns, 50µs pulse width +/-15ns |
Voltage Amplitude |
+/- 125V – 2kV into 50Ω, +/-240-4kV into 1000Ω |
Burst Duration |
15ms+/- 3ms@5kHz, .75ms +/- .15ms @100kHz |
Burst Period |
300ms +/- 60ms |
Frequency |
1-100 kHz, in .5k Hz steps, +/- 10% |
Combination Wave Surge: |
|
Standards |
IEC 61000-4-5 ed. 3, EN 61000-4-5, ANSI C62.41.2 Cat A& B, UL 1449 ed. 3 sec. 37, ANSI C62.45 |
Open Circuit Voltage Waveform |
1.2µs Front Time +/- 30%, 50µs Duration +/- 20% |
Short Circuit Current Waveform |
8µ Front Time +/- 20%, 20µs Duration +/- 20% |
Peak Open Circuit Voltage |
250V–6kV +/- 10% @2 Ω, 20.8-500A +/- 10%@12 Ω |
Peak Short Circuit Current |
125A–3kA +/-10%@2 Ω, 20.8-500A +/-10%@12Ω |
Pulse Repetition Rate |
2 pulses/min @6kV |
100 kHz Ringwave: |
|
Standards |
IEC 61000-4-12, EN 61000-4-12, ANSI C62.41.2 Cat A& B, ANSI C62.45 |
Open Circuit Voltage Waveform |
100kHz Damped Cosine, risetime=.5µs+/-30%, Damping= 40%/peak |
Peak Open Circuit Voltage |
250V–6kV +/- 10% @12 Ω or @ 30 Ω |
Peak Short Circuit Current |
200A +/-10%@30 Ωohms,500A +/-10%@12 Ω |
Pulse Repetition Rate |
Up to 2 pulses /min@6kV |
Telecom 10x700: |
Optional – can only have Ringwave or Telecom |
Standards |
IEC 61000-4-5, EN 61000-4-5, TIA-968-B, ITU k.44, k.45. |
Open Circuit Voltage Waveform |
10µs Front Time +/- 30%, 700µs Duration +/- 20% |
Short Circuit Current Waveform |
5µ Front Time +/- 20%, 320µs Duration +/- 20% |
Peak Open Circuit Voltage |
250V–6kV +/- 10% @15 Ω or @40 Ω |
Peak Short Circuit Current |
6.25-150A +/-10%@40 ohms – external 25Ω |
Repetition Rate |
1 impulse per min @ 6kV |
Dips and Interrupts: |
|
Standards |
IEC 61000-4-11, Ed.2 |
Voltage Dips: |
40%, 70%, 80% residual voltage +/-5% |
Short Interrupts |
0% residual voltage +/-5% |
Transition Time |
Risetime, falltime ≥1µs & ≤5µs |
Input Voltage/Current |
50-250VAC, 50/60Hz, 16A Max. |
Weight |
39kg, 85 pounds |
Dimensions |
23 x 43 x 65 cm, 8 x 17x26 inches |
All Accessories
Coupler | Capabilities |
CCL | Capacitive Coupling Clamp conforming to IEC 61000-4-4 |
CCLC |
Capacitive Coupling Clamp Cover |
PK1001D |
Differential High Voltage Probes |
CM-3CD-Surge/32 | Three phase, 32A External CDN for Surge Testing |
CM-I/OCD |
External 8-wire coupler/decoupler for I/O signal lines |
EFT-ATTN-CAL |
Eor EFT verification: Set of one 50Ω and one 1000Ω precision loads having >400MHz BW, one 20dB Attenuator with 1GHz BW |
CCL-VERIFY |
Manual 3-phase coupler for Surge (8 kV) only (combination and ring wave) with EUT current up to 100 A / 200 A |
Models/Standards
Models:
All of the Thermo Fisher EMC Pro equipment we carry are upgraded to the latest version of the associated standards and have the Plus distinction. This allows our customers the reassurance that the test generator provides the required waveforms. Below is a list of different models available through Thermo Keytek.
- PRO-BASE
- PRO-EFT
- PRO-SURGE
- PRO-TELECOM
- PRO-RING
- PRO-PQF
- PRO-HPOWER
- PRO-HPULSE
Standards:
Surge/Combination Wave
- IEC 61000-4-5 Ed. 3
- EN 61000-4-5
- ANSI C62.41.2 Cat A & B
- UL 1449 Ed. 3 Sect. 37
- ANSI C62.45
PQF/Dips
- IEC 61000-4-11
Ring Wave - Surge
- ANSI C62.41.2, Cat A, B
- IEC 61000-4-12
- EN 61000-4-12
- ANSI C62.45