Coupling NX7 bsr-3-690-63 - Rent 63Amp 3-Phase Surge & EFT CDN

EM Test 125A Automatic Coupler:

  • EUT Supply 3-Phase 690VAC/1kVDC 125Amps
  • Compliant Current Ratings <32 A, 63 A, 125 A
  • Must be Used with 400Volt NX5/NX7 Test Systems
  • Includes Laptop with Iec.control Software Installed. 
  • Automatic Selection of Combination Wave Impedance

Options/Equipment


Get weekly, two week, and monthly rentals on all products

Got Questions about the equipment? Get technical support

Two FREE days transit time with every rental

Request A Quote

Coupling Accessories

Description

Overview

The BSR-3-480-32 is an automated 3-phase coupler decoupler network (CDN) designed for electrical fast transients/burst and surge testing to IEC 61000-4-4 and IEC 61000-4-5. This equipment is limited to 3 x 480 VAC with a current rating of 32 Amps 50/60 Hz for the EUT. It provides coupling in accordance with EN/IEC 61000-4-4, EN/IEC 61000-4-5, and EN/IEC 61000-4-12 for burst up to 5.5kV, surge and ring wave to 7kV. This network can be rented with the EM Test NX7 series, or individually to supplement testing needs. This system comes standard with a current ISO 17025 accredited calibration and associated documentation. This CDN can be controlled via the immunity generator or through control software via PC.

Data SheetUser Manual

Specifications

Common Standards

EN/IEC 61000-4-4, EN/IEC 61000-4-5, EN/IEC 61000-4-12

Three-phase CDN

According to EN/IEC 61000-4-4, EN/IEC 61000-4-5 and -4-12 for Burst up to 5.5 kV Surge and Ring wave up to 7 kV

Current ranges*

<32 A, 63 A

Connectors

IN (AC): CEE 690/125,

IN (DC): MC Safety plug 6 mm,

OUT: MC Safety plug 6 mm

EUT supply   

3 x 690 V, 125 A, 50/60 Hz 1,000 VDC, 125 A

Dimensions        

Size: RAC 34 IEC included, Weight: approx. 250 kg

INTERFACES


Sys.link

26 pin high density connector to control an external devices

Connector EUT supply    

3 x 480 V, CEE plug, 16 A to 125 A,

3 x 690 V, CEE plug, 16 A to 63 A,

plugs for all other current ratings

Connector to EUT

up to 32 A: 4 mm safety plug,

63 A: 6 mm safety plug,

125 A: 6 mm safety plug,

200 A: 10 mm safety plug

MAINS (CONTROL)


Supply voltage       

85 V - 264 V

Frequency

50/60 Hz

Power

approx. 75 W Standby: 100 mA

Fuses           

115 V: 2 x 4 A slow blow,

230 V: 2 x 2 A slow blow

SAFETY


Safety standard

IEC/EN 61010

Security circuit     

Control input (24 VDC)

Warning lamp

Floating contact (max. 60 V/2 A)

ENVIRONMENT


Temperature           

10 °C to 35 °C

Humidity

30 % to 75 %, non condensing

Atmospheric pressure

86 kPa (860 mbar) to 106 kPa (1,060 mbar