Description
Overview
The CLEM-6146 Electromagnetic, or ‘EM’ Clamp is designed for use as a clamp injection device for conducted immunity tests in the frequency range of 150 kHz to 230 MHz according to IEC/EN 61000-4-6. The clamp can also be used for tests over its extended frequency range of 10 kHz to 1 GHz.
The cable connected between the Equipment Under Test (EUT) port to be tested and its Auxiliary Equipment (AE) is placed into the EM Clamp. The EM Clamp provides a means by which RF energy is coupled onto the cable and decoupled from the AE path, so that the energy is directed toward the EUT.
The CLEM-6146 incorporates split-core ferrites which allows the top half of the hinged enclosure to open so that the cable under test can easily be laid into position. Once closed, rotating the two knobs on the front of the clamp onequarter turn engages the locking mechanism which compresses the top and bottom halves of the ferrites together around the cable.
Data SheetUser ManualSpecifications
Frequency Range |
150 kHz to 230 MHz |
Extended Frequency Range |
10 kHz to 1 GHz |
Nominal Impedance |
50Ω |
Input Connector |
Type-N (female) |
Maximum Cable Diameter |
0.75” (19 mm) |
Maximum Input Power 10 kHz to 100 MHz 100 MHz to 230 MHz 230 MHz to 1 GHz |
100 Watts for 15 minutes 100 Watts for 3 minutes 50 Watts for 3 minutes |
Dimensions (L) x (W) x (H) |
25.6” x 4.9” x 4.9” (650.2 mm x 124.5 mm x 124.5 mm) |
Weight
|
13 lbs (5.9 kg) |