Associated Research Hypot 3870 - Rent 5/6kV AC/DC Hipot Tester

Availability:
In Stock - Order By 3PM Ships Today!

Hipot & Safety Test System:

  • 5kVA/6kVA AC/DC Hipot, IR, & Ground Continuity 
  • Compact & Easy to Use Bench Top Unit
  • Available TV-B 2 Verification Box Rentals
  • 2 AC & 3 DC Current Range Options (+/-2%) 
  • Current ISO 17025 Cal & 2 Days Free Transit Time

Price:
Sale price$300
Request A Quote

Description

Hipot & Ground Tester Features 

      • 5kV AC, 6kV DC, Hipot, Ground, & Insulation Tester
      • Compact & Easy to Use Bench Top Unit
      • Ideal for Dielectric Withstand & Insulation Tests
      • Includes Capabilities of Hypot 3805 & 3865 series
      • Free Technical Support & 2 Days Free Transit 

 

Overview

The new Hypot® Series raises the bar for safety and efficiency in production line Hipot Testers. Whether your application requires AC Withstand tests, DC Withstand tests, or Insulation Resistance tests, The Hypot Series Hipot Tester provides innovations that will allow you to begin achieving best practices for safety and efficiency without the need for an expensive automated test system or connection to a PC. Stop handwriting thousands of test results and improve traceability with on-board data storage and data transfer capabilities.

Automate your production line with direct barcode connection, which allows you to quickly associate products with pre-programed test files. Keep operators safe with SmartGFI®, Safety Interlock, and PPE compatible accessories. You can even connect the Hypot® Series with the HYAMP® Series to form a seamless safety compliance system for Hipot and Ground Bond Testing. The Hypot Series includes the following models: 3805, 3865, 3870.

Data SheetUser Manual Quick Start Guide

Specifications

Input Specifications


Voltage 

 100 – 120 VAC / 200 – 240 VAC ± 10% Auto Range

Frequency

50/60 Hz ± 5%

Fuse

3.15 A, Fast Blow 250 VAC


Dielectric Withstand Test Mode


Output Rating

5 kVA @ 20 mAAC 6 kVA @ 7.5 mADC (3865/3870 only)

Maximum Limit

AC

Range: 0.00 – 20.00 mA  Resolution: 0.01 mA

DC

Range:  0 – 7500 µA Resolution: 1 µA Accuracy: AC and DC ± (2% of setting + 2 counts)

Minimum Limit

AC 

Range:  0.000 – 9.999 mA Resolution: 0.001 mA

DC 

Range:  0.0 – 999.9 µA Resolution: 0.1µA Accuracy: AC and DC ± (2% of setting + 2 counts)

Frequency

Range: 1-9, ON/OFF Select

Ground Fault Interrupt

 GFI Trip Current: 450 µA max (AC or DC), Fixed

HV Shut Down Speed: < 1 msec

Current Display

AC

Range 1: 0.000 – 4.000 mA

Range 2: 3.50 – 20.00 mA

DC

Range 1:  0.0 µA – 400.0 µA

Range 2: 0.350 mA – 4.000 mA

Range 3: 3.50 mA – 7.50 mA Accuracy: All Ranges ± (2% of reading + 2 counts)

 DC Output Ripple

≤ 5% Ripple rms at 6 kVDC @ 7.5 mA Resistive Load

RAMP-HI Selectable

Range: 0.0 – 7,500 µA, User Selectable

Charge-LO

0 – 350 µA DC or Auto Set

Peak Short Circuit Current

< 50 msec for no load, < 100 msec for capacitive load

The maximum capacitive load vs. output voltage:

1µF < 1KV 0.08µF < 4KV 0.75µF < 2KV 0.04µF < 5KV 0.5µF < 3KV 0.015uF < 6KV

Insulation Resistance Test Mod

 

Voltage Setting 

Range: 30 – 1,000 VDC 

Resolution: 1 V 

Accuracy: ±(2% of setting + 5 V)

Resistance Display

Range: 1 – 50,000 MΩ