Associated Research HypotULTRA 7804 - Rent 4-in-1 Tester

AC/DC Hipot & Ground Tester

Associated Research HypotULTRA 7804 - Rent 4-in-1 Tester

Availability: In Stock Shortly

HypotULTRA Tester Features 

      • 5kV AC, 6kV DC, Hipot, Ground, & Insulation Tester
      • Touch Screen & Easy to use Menus
      • Ideal for Dielectric Withstand & Insulation Tests
      • Capabilities of HypotULTRA 7820 & 7850
      • Free Technical Support & 2 Days Free Transit 

 

Overview

Our new HypotULTRA® models provide all the tools you need to modernize your production line with best-in-class 4-in-1 test capability and a slim 2U design. We’ve added 40A AC Ground Bond test capability to HypotULTRA®’s already impressive feature list for manufacturers that aim to adopt best testing practices without sacrifi cing productivity. 

Whether you’re looking to improve traceability with on-board data storage, increase effi ciency with our intuitive touch screen interface and direct barcode scanner connection, or automate with a variety of communication interfaces, HypotULTRA® was designed to take your production line to the next level.

$250.00

HypotULTRA Tester Features 

      • 5kV AC, 6kV DC, Hipot, Ground, & Insulation Tester
      • Touch Screen & Easy to use Menus
      • Ideal for Dielectric Withstand & Insulation Tests
      • Capabilities of HypotULTRA 7820 & 7850
      • Free Technical Support & 2 Days Free Transit 

 

Overview

Our new HypotULTRA® models provide all the tools you need to modernize your production line with best-in-class 4-in-1 test capability and a slim 2U design. We’ve added 40A AC Ground Bond test capability to HypotULTRA®’s already impressive feature list for manufacturers that aim to adopt best testing practices without sacrifi cing productivity. 

Whether you’re looking to improve traceability with on-board data storage, increase effi ciency with our intuitive touch screen interface and direct barcode scanner connection, or automate with a variety of communication interfaces, HypotULTRA® was designed to take your production line to the next level.

Data SheetUser Manual Quick Start Guide
Input Specifications


Voltage 

 100 – 120 VAC / 200 – 240 VAC ± 10% Auto Range

Frequency

50/60 Hz ± 5%

Fuse

3.15 A, Fast Blow 250 VAC


Dielectric Withstand Test Mode


Output Rating

5 kVA @ 20 mAAC 6 kVA @ 7.5 mADC (3865/3870 only)

Maximum Limit

AC

Range: 0.00 – 20.00 mA  Resolution: 0.01 mA

DC

Range:  0 – 7500 µA Resolution: 1 µA Accuracy: AC and DC ± (2% of setting + 2 counts)

Minimum Limit

AC 

Range:  0.000 – 9.999 mA Resolution: 0.001 mA

DC 

Range:  0.0 – 999.9 µA Resolution: 0.1µA Accuracy: AC and DC ± (2% of setting + 2 counts)

Frequency

Range: 1-9, ON/OFF Select

Ground Fault Interrupt

 GFI Trip Current: 450 µA max (AC or DC), Fixed

HV Shut Down Speed: < 1 msec

Current Display

AC

Range 1: 0.000 – 4.000 mA

Range 2: 3.50 – 20.00 mA

DC

Range 1:  0.0 µA – 400.0 µA

Range 2: 0.350 mA – 4.000 mA

Range 3: 3.50 mA – 7.50 mA Accuracy: All Ranges ± (2% of reading + 2 counts)

 DC Output Ripple

≤ 5% Ripple rms at 6 kVDC @ 7.5 mA Resistive Load

RAMP-HI Selectable

Range: 0.0 – 7,500 µA, User Selectable

Charge-LO

0 – 350 µA DC or Auto Set

Peak Short Circuit Current

< 50 msec for no load, < 100 msec for capacitive load

The maximum capacitive load vs. output voltage:

1µF < 1KV 0.08µF < 4KV 0.75µF < 2KV 0.04µF < 5KV 0.5µF < 3KV 0.015uF < 6KV

Insulation Resistance Test Mod

 

Voltage Setting 

Range: 30 – 1,000 VDC 

Resolution: 1 V 

Accuracy: ±(2% of setting + 5 V)

Resistance Display

Range: 1 – 50,000 MΩ