Chroma’s Guardian 19056/19057 Hipot Analyzer is specially designed for testing and analyzing ultra-high withstand voltage. The series of models include 10kVac/12kVdc/20kVdc with maximum AC20mA/DC10mA output that can perform AC/DC withstand voltage and insulation resistance tests with contact check during production line test. In addition to the patented OSC (Open Short Check), High Voltage Contact Check is added to test the components with high insulation capability when high voltage outputs to improve the testing reliability and efficiency.

Data SheetUser Manual Short Check Function

Specifications

Input Specifications

 19057-20 DCW

Output Voltage

DC : 0.1 ~ 20kV

Load Regulation

± (1% of output + 10V), Rated load

Voltage Accuracy

± (1.5% of reading + 0.1% of full scale)


Cutoff Current

0.0001~5 mA

Current Resolution

DC : 0.1µA

Test/Ramp/Fall/Dwell Time

0.3 ~ 999 sec., continue / 0.1 ~ 999 sec., off / 0.1 ~ 999 sec., off / 0.1 ~ 999 sec., off

Insulation Resistance Test

Output Voltage

DC : 0.1 ~ 5kV

Voltage Resolution

2V

Voltage Accuracy

1.5% of setting + 0.5% of full scale

IR Range

0.1MΩ ~ 50GΩ

Resistance Resolution

0.1MΩ

Resistance Accuracy ≧ 0.5kV

1MΩ ~ 1GΩ : ± 3% of reading + 0.5% of full range

1GΩ ~ 10GΩ : ± 5% of reading + 1% of full range

10GΩ ~ 50GΩ : ± 10% of reading + 1% of full range

Resistance Accuracy <0.5kV

1MΩ ~ 1GΩ : ± 5% of reading + (0.5*500/Vs)% of full scale

Flashover Detection

 

Setting Mode

Programmable setting

Detection Current

DC : 1mA~10mA