Description
Overview
The MegaPulse Debif D5-5P by Compliance West tests to 5,000 volts including direct energy readout (J) for energy reduction tests, built-in pass/fail reference. Our EMC test equipment rentals come with an A2LA accredited calibration and technical support. This medical based immunity testing equipment displays selected output voltage on digital front panel. This system meets a variety of defibrillation simulation requirements for a diverse set of applications.
The Immunity Generator defibrillator proof test set to The Mega Pulse DefibD5-PF offers a USB computer connection for remote operation. This allows for increased flexibility in setups, and offers more robust and problem resistant connection. On the test equipment side, the interface is RS-232, which offers a stronger, more error-free protocol for more trouble-free operation.
This Compliance West medical based product is designed to meet IEC 60601-1 figure 10, IEC60601-1 figure 11, IEC 60601-2-34,2-49; EC-13, EC-53 and many others. This immunity product provides testing solutions for many standards that require 5000v with different output wave shaping.
Data SheetUser ManualSpecifications
Common Standards | IEC 60601-1:2005 figure 10; IEC60601-1:2005 figure 11; IEC 60601-2-34,2-49; EC-13 and EC-53 |
Voltage Source | 0-5100 V ± 1% |
Main Capacitance | 32µF ± 5% |
Inductances | 500µH and 25µH |
Duty Cycle | 20s between pulses, ± 1% 100Ω resistor bank, up to 70 pulses
12s between pulses, +1%<-5% 100Ω resistor bank, continuous |
Energy Measurement | Works only when 25mH inductor selected
Display Resolution: 1 Joule Repeatability: ± 1 Joule between pulses |
Dimensions | 17”x18”x17” |
Weight | 40 lbs |
Packing List/Accessories
Included with Rental
Below you will find a list of common accessories that you can expect to be delivered with the equipment. This is a generic standard list and should you require an exact connector or accessory please contact us prior to order shipping. We are also excited to offer many different testing accessories not standard with the immunity generator.
MegaPulse Defib D5-PF
- Pulse Generator
- Printed Manual
- Calibration Documentation
- (3) HV Test Leads (Red)
- (2) HV Test Leads (Black)
- Software CD
- USB Box
- DB-9 Cable (Male to Male)
- Case
Standards
Widely Used Standards
This list is a more comprehensive list of all the standards covered with this immunity generator. This is by no means all the standards covered, but the major ones from the largest manufacturers and most widely used standards.
- IEC 60601-1:2005 (Issue 3) Figure 9 and 10 Defib-proof Differential- and Common-Mode Test (IEC 60601-1 Figure 50 in older versions).
- IEC 60601-1:2006 Figure 11 Energy Reduction Test;
- IEC 60601-2-4: 2010 Figure 201.110 Overload Test (optional)
- IEC 60601-2-34 Issue 2 (optional);
- IEC 60601-2-34 Issue 3 (included);
- IEC 60601-2-49;
- EC-13;
- EC-53;
- IEC 60601-2-25:2011;
- IEC 60601-2-27:2011.
Options
Common Options
Item | Capabilities |
00-MP-ESTOP-01 | Emergency Stop Option (When Activated Isolates Outputs and Discharges Main Capacitor. Option adds 3U level) |
00-MP-BB-001 |
(Medical Breakout Box. The Breakout Box is a separate box) |
00-MP-RC-01 |
(MegaPulse Remote Control 01) |
00-MP-RC-02 | (MegaPulse Remote Control 02) |
00-MP-RC-01-CA-20 |
(MegaPulse Remote Control Cable 20ft) |
00-SG |
(Sig. Gen. for IEC 60601-2-27, 10Hz 20Vp-p) |
00-34 |
(Testing to IEC 60601-2-34, Invasive Blood Pressure Test, Ed. 2.0 only, adds 3U level) |
00-2-4 |
00-2-4 (Testing to IEC 60601-2-4:2010, Defib Overload Test, 10 ohm output, adds 3U level) |
00-GVAC |
(Gigavac Vacuum Relay Switch for D5-PF Trigger) |
00-CASE-6U |
(Pelican Case with Foam Interior for D5-PF without option 00-34 or 00-2-4) |
A2LA-TECH-MP |
(ISO 17025 Calibration by Techmaster, adds 2 week to delivery time) |