ECAT system for Surge, with 3 phase coupler/decoupler for EN IEC 61000-4-5, ANSI C62.41, C62.45



ECAT system for Surge, with 3 phase coupler/decoupler for EN IEC 61000-4-5, ANSI C62.41, C62.45

 

Controller-E103

Controller for the ECAT system with front panel and software operation.

Combination Wave Surge-E501B:

 

  Standards

IEC 61000-4-5 ed. 3, EN 61000-4-5, ANSI/IEEE C62.41.2 Cat. B, UL 1449 at 3kA

  Open Circuit Voltage Waveform

1.2µs, Front Time ± 30%, 50µs Duration ±20%

  Short Circuit Current Waveform

8µs Front Time ±20%, 20µs Duration ±20%

  Peak Open Circuit Voltage

250V–6kV ± 10% @2 Ω,  250-6kV ±10% @12 Ω

  Peak Short Circuit Current

125A–3kA ±10%@2 Ω, 20.8-500A ±10%@12Ω

  Pulse Repetition Rate

2 pulses/min. @6kV

100 kHz Ringwave-E503:

 

  Standards

ANSI/IEEE C62.41.2 Cat. A& B, UL 864, EN IEC 61000-4-2, ANSI C63.16

  Open Circuit Voltage Waveform

100kHz Damped Cosine, Rise Time=.5µs±30%,

Damping= 40%/peak

  Peak Open Circuit Voltage

200V–6.6kV ± 10% @12 Ω or @ 30 Ω, Front Time0.5µs ±30%

  Peak Short Circuit Current

200A ±10%@30 Ω, 500A±10%@12 Ω

  Pulse Repetition Rate

6 pulses/min.

  Line Sink Accuracy

± 15° with optional coupler/decoupler

Integrated CDN-E4554:

 

  AC Voltage and Current              

50-480 VAC, 32A, 5 lines/3 phases

  DC Voltage and Current

N/A