NSG 4070 110 Watt BCI System - 10khz-400MHz Turnkey Equipment

110 Watt Current Injection System

NSG 4070 110 Watt BCI System - 10khz-400MHz Turnkey Equipment

Conducted Immunity - BCI System 

      • Bulk Current Injection (BCI) Test System
      • Laptop & ISO 17025 Calibration Included
      • Integrated signal generator 9 kHz - 1 GHz
      • External 110 Watt Amplifier 10kHz-400MHz
      • Free Technical Support & 2 Days Free Transit
      • Includes Fisher F-120-6A 10kHz-400MHz CIP
      • Turnkey Solution - Arrives Ready for Testing!

Overview

The bulk current injection system includes the Teseq NSG4070, Laptop with software, F-120-6A current injection probe (CIP), CBA 400M-110 and associated cabling calibration fixtures for 10kHz-400MHz. This system is ideal for higher requirements of IEC 61000-4-6, ISO 11452-4, MIL-STD-461 CS114 and many others. The Teseq external amplifier can be rented individually or with the system to offer a BCI test solution.  Take advantage of ISO 17025 calibrations, a variety of accessories, a quick delivery on this equipment.

 

Teseq NSG 4070 RF Test Accessories

Conducted Immunity - BCI System 

      • Bulk Current Injection (BCI) Test System
      • Laptop & ISO 17025 Calibration Included
      • Integrated signal generator 9 kHz - 1 GHz
      • External 110 Watt Amplifier 10kHz-400MHz
      • Free Technical Support & 2 Days Free Transit
      • Includes Fisher F-120-6A 10kHz-400MHz CIP
      • Turnkey Solution - Arrives Ready for Testing!

Overview

The bulk current injection system includes the Teseq NSG4070, Laptop with software, F-120-6A current injection probe (CIP), CBA 400M-110 and associated cabling calibration fixtures for 10kHz-400MHz. This system is ideal for higher requirements of IEC 61000-4-6, ISO 11452-4, MIL-STD-461 CS114 and many others. The Teseq external amplifier can be rented individually or with the system to offer a BCI test solution.  Take advantage of ISO 17025 calibrations, a variety of accessories, a quick delivery on this equipment.

 

Data Sheet - 4070Teseq - Software DownloadData Sheet - F120-6A Fischer

Common Standards

EN/IEC 61000-4-6, EN/IEC 60601-1-2, ISO 11452-4, EN/IEC 62132

Generator

 

 Frequency Range

9 kHz to 1 GHz

 Resolution

1 Hz

 RF Level Range

-60 dBm to +10 dBm

 RF Level Resolution

.1 dBm


 Amplitude Modulation Frequency

1 Hz to 50 kHz

 Amplitude Modulation Resolution

1 Hz

 Pulse Modulation Rise/Fall Time

<1µs (10% to 90%)

 Pulse Modulation Frequency Range

1 Hz to 50 kHz

 Pulse Modulation Freq. Resolution

1 Hz

 Pulse Modulation Duty Cycle

10% to 90%

Power Meter

 

  Frequency Range

9 kHz to 1 GHz

  Line Measurement Range

Channel 1 -35 dBm to +27 dBm, Ch 2&3 -45 dBm to 20dBm

  Maximum Input

+28 dBm

  Noise Level

>5 dB below the measurement range

  Input Return Loss

>20 dB (below 500MHz), >17 dB (500MHz to 1 Ghz)

  Connector

BNC socket, 50Ω

Power Amplifier - Built In

 

  Output Power

Max 80W

  Frequency Range

150 kHz to 230 MHz

  Input Impedance

50 ohm

  Output Impedance

50 ohm

  Direct RF Output

To control an external amplifier

  Input Return Loss

Min. 10 dB

  Output Return Loss

Nominal Min. 9.5 dB, 0 dB without damage

  Gain

Min 50 dBm

  Linear Output Power

Min 48 dBm

  Max. Input Power without Damage

Max. +10 dBm

  Harmonic Distortion

Typ. <-20 dBc

Weight

15 kg, 33 pounds

Dimensions

45x43x15 cm, 17x16x6  inches

 Frequency Range (Instantaneous)

 10 kHz to 400 MHz

 Rated Output Power

 110 W Minimum

 Output Power at 1dB Gain Compression

 80 W Minimum (110 W typical below 100 MHz)
 Frequency Range (Instantaneous)

 10 kHz to 400 MHz

 Rated Output Power

 110 W Minimum

 Output Power at 1dB Gain Compression

 80 W Minimum (110 W typical below 100 MHz)

 

Included with Rental of Conducted RF Test System

Below you will find a list of common accessories that you can expect to be delivered with the equipment. This is a generic standard list and should you require an exact connector or accessory please contact us prior to order shipping. We are also excited to offer many different testing accessories not standard with the immunity generator. This does not include any coupler/decoupler items or accessories. A detailed list of these can be found on the "All Accessories" tab. 

 

NSG 4070 (C) Test Accessories 

  • NSG 4070
  • Power Cable
  • Printed Manual
  • Calibration Documentation
  • Pasternak PE 7021-6 6dB Attenuator 
  • Pasternak 20db Attenuator 
  • N Female to N Female adapter
  • N Female to BNC Female Adapter
  • N Male to BNC Female Adapter
  • 50 Ohm Termination Load
  • USB Cable 
  • 9 Pin D Shaped Serial Cable
  • N to N Cable LE 239
  • N to N Cable 12inch
  • Silver U Shaped Jumper
  • BNC to BNC Cable
  • USB Stick with Software
  • Ethernet Cable
  • USO USB Box
  • USO 10m. Cable
  • N to BNC Cable
  • Case

 

   Common BCI accessories

  • FCC-BCICF-1 Calibration Fixture
  • ATT 3dB 100W attenuator
  • F40
  • F55
  • ZA-41 20dB (2 pcs)
  • ZA-04 50Ohm termination Load
  • N-N 25 CM
  • N-N 1 Meter
  • BNC to BNC 50cm
  • BNC-BNC 2 Meter
  • N to BNC various adapters
  • Calibration Data USB
  • Printed Manual
  • Case

    Widely Used Standards

     

    This list is a more comprehensive list of all the standards covered with this conducted RF Generator. For assured compliance to a particular standard, confirm prior to rental. This conducted immunity generator has the ability to either cover all, or part of the standards below.

     

    • IEC/EN 61000-4-3
    • IEC/EN 61000-4-6
    • IEC/EN 61000-4-20
    • IEC/EN 61000-4-21
    • CISPR 24
    • EN 61000-6-1
    • EN 61000-6-2
    • IEC 60533
    • IEC 60601-1-2
    • IEC 61326
    • MIL STD 461
    • RTCA/DO 160
    • ISO 11452-2
    • ISO 11452-3
    • ISO 11452-4
    • ISO 11452-5
    • SAE J1113-4
    • SAE J1113-24

    Rent a variety of Conducted RF Immunity accessories with test solution.

    Take advantage of A2LA ISO 17025 calibrations and fast deliveries!

     

    Conducted RF Probes, Clamps, & Devices
    Item Capabilities
    EM101 EM Clamp .15 to 1,000 MHz, 100 W max, incl: FTC 101 decoupling clamp
    F120-9A Clamp-on injection probe, 10kHz to 230MHz, 125 CW, calibration fixture
    CDN-AF4 100v, 1.5amp. coupler for 4 unshielded lines
    CDN-S15 100V, 1.5 amp. CDN for 15 shielded lines
    CDN-M2/3/16A 240VAC, 16A coupler for 2 or 3 power lines
    CDN-M5/32A

    400VAC, 32A coupler for 5 power lines

    Additional Available through Manufacturer

    MD 4070A Monitoring Device 4kHz - 400MHz
    SW 4070, SW 4070-MAN RF Switch NSG 4070 up to 2GHz
    ATN Series 6 dB Attenuators

    Built in Amplifiers NSG 4070

    There are a variety of different amplifier options that can be built into this conducted and radiated test solution. All of these Teseq amplifier built in these products are class A power amplifiers. These range from 260 Watts at the highest to 35 Watts at the lowest. All rental solutions from Transients Specialists come with either 75 Watt or 80 Watt built in amplifiers covering the 150kHz to 230Mhz frequency range for this generator. Below you will find the list of the amplifier watt choices and corresponding frequency ranges.

     

    • 260 Watt Amplifier - 10 kHz (4 kHz) to 400 MHz 
    • 110 Watt Amplifier - 10 kHz (4 kHz) to 400 MHz 
    • 80 Watt Amplifier - 150 kHz to 230 MHz
    • 60 Watt Amplifier - 10 kHz (4 kHz) to 400 MHz 
    • 45 Watt Amplifier - 9 kHz to 1 GHz 
    • 40 Watt Amplifier - 10 kHz to 400 MHz 
    • 35 Watt Amplifier - 150 kHz to 230 MHz

     

    NSG 4070 - Teseq Multifunction Generator 

     

    Description: This video made by Teseq, now owned by Ametek, provides an overview of this conducted RF immunity test system. It provides an overview of the system itself, accessories, including coupling clamps, controls, interface, and manipulating testing parameters. It also mentions briefly about compliance to IEC 61000-4-6 and other testing standards.