UCS500n7 - EM Test EFT to 5.5kV, Surge and 100khz Ring to 7kV, Built-in Single Phase Power Coupler

RENTAL

UCS500n7 - EM Test EFT to 5.5kV, Surge and 100khz Ring to 7kV, Built-in Single Phase Power Coupler

 

EMTest EMC Immunity Tester for  EFT to 5.5kV and Surge and 10khz ringwave to 7kV with built-in single phase coupler and software, optional Dips and Interrupts testing with the V4780 S2 external variac



 

EMTest EMC Immunity Tester for  EFT to 5.5kV and Surge and 10khz ringwave to 7kV with built-in single phase coupler and software, optional Dips and Interrupts testing with the V4780 S2 external variac

Common Standards

EN IEC 61000-4-4, EN IEC 6100-4-5, EN IEC 61000-4-12, EN IEC 61000-4-11, CEMark, ANSI IEEE C62.41 cat. B

EN 61000-6-1, -6-2

Integrated CDN:

 

  AC Voltage and Current              

300V/16A; 50/60Hz

  DC Voltage and Current

300V/10A

EFT Burst 5/50ns:

 

  Standards

IEC 61000-4-5, ed.3, CEMark

  Voltage Waveform into 50Ω load

5ns Rise Time ±1.5ns,  50µs pulse width ±15ns

  Voltage Waveform into 1000Ω load

5ns Rise Time ±1.5ns,  50µs pulse width -15/+100ns

  Voltage Amplitude

200V-5500 V ±10%, 100V-2,750V ±10% into 50Ω

  Burst Duration  

td = 0.10ms - 999ms

  Burst Period

tr = 10ms - 9,999ms

  Frequency

f = 0.1kHz - 1,000kHz

Combination Wave Surge:

 

  Standards

IEC 61000-4-5 ed. 3, EN 61000-4-5, ANSI C62.41.2 Cat A& B, UL 1449 ed. 3 sec. 37, ANSI C62.45, EN 61000-6-1, -6-2

  Open Circuit Voltage Waveform

1.2µs Front Time ±30%,  50µs Duration ± 20%

  Short Circuit Current Waveform

8µ Front Time ±20%, 20µs Duration ± 20%

  Peak Open Circuit Voltage

250V–7kV ±10% @2 Ω

  Peak Short Circuit Current

125A–3.5kA  ±10%@2 Ω

  Pulse Repetition Rate

2s – 999s

100 kHz Ringwave:

 

  Standards

IEC 61000-4-12, EN 61000-4-12, ANSI C62.41.2 Cat A& B,   ANSI C62.45

  Open Circuit Voltage Waveform

100kHz Damped Cosine, Rise Time=.5µs±-30%,

Damping= 40%/peak

  Peak Open Circuit Voltage

250V–6kV ± 10% @12 Ω or @ 30 Ω

  Peak Short Circuit Current

200A±10%@30 Ωohms, 500A ±10%@12 Ω

  Pulse Repetition Rate

1s – 999s

Dips and Interrupts:

Optional External V4780 S2

  Standards  

EN/IEC 61000-4-11 and EN 61000-6-1, -6-2

  Voltage Dips:

40%, 70%, 80% , 100%residual voltage ±5%

  Short Interrupts

0% residual voltage ±5%

  Transition Time

Rise Time, Fall Time ≥1µs & ≤5µs

  Input Voltage/Current

50-250VAC, 50/60Hz, 16A Max.

Weight

29kg, 64 pounds

Dimensions

43x43x48cm, 17x17x19 inches

I would recommend Transient Specialists. Transient Specialists always provides quality products within the established timeframes. The equipment always meets the specifications and properly shipped within the time frames within the time frames promised. 

Scott D., TUV SUD Canada

Transient Specialists has been outstanding at working with us for our specialized EMC equipment rental needs. Fast shipments and excellent accommodation for some customized setups. I recommend them highly and they are a pleasure to partner with!

John H., Gates Air, Inc.

Laura, I want to thank you and Goran for your excellent customer service and turn around time on the rentals and the calibration documents.  

Adam P., Clark Testing

Working for a testing company I deal with renting equipment all the time.  Transient Specialist is the first company I call on my list of rental houses because of their great personnel, prompt delivery of the equipment, knowledgeable and accommodating to the customers needs.

Richard T., National Technical Systems