EMC Partner IMU-MGS
Included Accessories
Can't find something? Contact UsRelated Equipment
Description
The EMC Partner IMU-MGS is a compact, multifunction EMC immunity test system designed for performing electrical transient testing on commercial and industrial electronic equipment. Combining multiple immunity test functions into a single platform, the IMU-MGS is well suited for EMC laboratories, product development teams, and manufacturers conducting both pre-compliance and compliance testing. Depending on the system configuration, the IMU-MGS can perform surge testing to IEC 61000-4-5 and electrical fast transient (EFT)/burst testing to IEC 61000-4-4, along with additional immunity tests when equipped with the appropriate modules and accessories.
The system incorporates an internal coupling/decoupling network (CDN), allowing surge and EFT disturbances to be coupled directly onto compatible single-phase AC or DC power lines. External coupling networks can also be added for testing three-phase equipment or EUTs with higher power requirements.
Designed to simplify transient immunity testing, the IMU-MGS provides programmable test parameters, automated test sequences, and software-based system control. Compatible accessories such as three-phase CDNs and EFT capacitive coupling clamps further expand its testing capabilities. Its combination of multiple transient functions, flexible configurations, and straightforward operation makes the EMC Partner IMU-MGS a versatile solution for laboratories performing a wide range of EMC immunity evaluations.
- IMU-MGS Mainframe
- EUT power input 1 (CDN)
- max. AC/DC 300V, 16A (fused 16A) with P module
- EUT power input 2 (CDN)
- max. AC/DC 300V, 16A (not fused) with P module
- Internal CDN freq. range
- DC, 50 Hz, 60 Hz
- Power freq. synchr.
- 50/60 Hz
- Coupling Burst
- L, N, PE, L+N, L+PE, N+PE, L+N+PE, direct out
- Coupling Surge
- 2 ?: L-N, direct out, 12 ?: L-PE, N-PE
- IMU-MGS Control Features
- Operating system
- EPOS propietary firmware
- Languages
- 8 menu languages, selectable
- User interface
- 7 inch capacitive touch display
- Connectivity
- gigabit ethernet, USB, RS485
- Environment meas.
- built-in temperature and humidity sensor
- EUT power monitor
- supply voltage, current and freq. on screen
- EUT supply voltage waveform
- monitoring BNC output connector max. 15 V
- EUT supply current waveform
- monitoring BNC output connector max. 15 V
- Surge monitoring
- peak voltage and current on screen, waveform available at BNC outputs
- Surge voltage waveform
- monitoring BNC output connector max. 15 V
- Surge current waveform
- monitoring BNC output connector max. 15 V
- Trigger out
- BNC connector, max. 6 V
- Trigger in
- auto, manual or via programmable BNC
- Programmable connectors
- 2 BNC connectors (inputs/outputs) as follows
- Programmable input functions
- Trigger Input, Start Test, Stop Test, EUT Fail, EUT Mark, Emergency Stop
- Programmable input max. voltage
- low range: 0 - 1.5 V, high range: 2.3 - 24 V
- Programmable output functions
- Running State, Safety Circuit State
- Programmable output max. U, I
- max. 24 V, max. 300 mA
- Synchronization source
- EUT power, direct out, external
- Synchronization angle
- 0 - 359 degree +/- 5 degrees, 1 degree step
- Input EUT power selection
- PWR1 and PWR2 selectable on the screen
- Power ON/OFF process
- sophisticated functions allow to switch EUT power ON/OFF at selectable voltage angle (0 - 359 degrees, 1 degree step)
- Impulse polarity
- positive, negative, alternating
- Automatic ramps
- test level, sync. angle, coupling path
- EUT overcurrent protection
- automatic programmable fuse up to 16 A
- PS3 control
- DC, 16.7 Hz, 50 Hz, 60 Hz supply power
- Safety features (standard)
- Emergency stop button on front panel red/yellow as per IEC 60947-5-5, IEC 60204-1, ISO 13850, Safety circuit
- IMU-MGS module: EXT-IMU4000 F4
- Standard
- IEC 61000-4-4 latest edition
- Output impedance
- 50 Ohm
- Voltage OC
- 0.2 kV - 4.4 kV
- Voltage into 50 ?
- 0.1 kV - 2.2 kV +/- 10 %
- Waveform into 50 ?
- 5 +/- 1.5 ns / 50 +/- 15 ns
- Voltage into 1 k?
- 0.191 kV - 4.18 kV +/- 20%
- Waveform into 1 k?
- 5 +/- 1.5 ns / 50 ns (-15/+100 ns)
- Spike frequency
- 1 kHz - 1 MHz
- Maximum spikes / second
- 8000 @ 1 kV
- Burst duration
- 0.01 ms - 30 ms, continuous up to 8 kHz
- Burst repetition
- 1 - 1000 ms
- Polarity
- positive, negative, alternating
- Synchronization
- 0 - 359 degrees, step 1 degree
- Ramps
- voltage, spike frequency, burst duration
- IMU-MGS module: EXT-IMU4000 S4
- Standard
- IEC 61000-4-5 latest edition
- Output impedance
- 2 ?
- Voltage OC
- 0.25 kV - 4.1 kV +/- 10 %
- Voltage waveform
- 1.2 us +/- 30 % / 50 us +/- 20 %
- Current SC
- 0.125 kA - 2.05 kA +/- 10 %
- Current waveform
- 8 us +/- 20 % / 20 us +/- 20 %
- Pulse repetition
- up to 1 / s @ 0.8 kV, 1 / 5 s @ 4.1 kV
- Polarity
- positive, negative, alternating
- Synchronization
- 0 - 359 degrees, step 1 degree
- Ramps
- voltage, synchronisation angle
- Magnetic pulse test
- IEC 61000-4-9, see MF1000-x coils
- IMU-MGS mainframe:
- Operating voltage
- 100 - 240 V (50/60 Hz) +/- 10 %
- Power consumption
- ON < 150 VA, standby < 15 VA
- Weight
- 16 kg (weight of modules not included)
- W x d x h
- 520 x 433 x 180 mm
- Version
- 19 inch unit, 4 UH