Teseq KEMZ 801A
Related Equipment
Description
IEC 61000-4-6 (EN 61000-4-6) defines three basic types of transducers used for the injection of test signals into the equipment under test (EUT). The basic requirement is to inject a known level of RF signal into the cable of the EUT at each test frequency and to determine a possible susceptibility of the EUT. The EM clamp injects the disturbance signal by a combination of inductive and capacitive coupling.
The normative Annex A of IEC/EN 61000-4-6 specifies the EM clamp and provides additional information regarding the clamp characterization. The KEMZ 801A complies to the standard. The adapter kit CAL 801A, which allows test level setting and correction factor measurement, is offered optionally. The calibration fixture CAL KEMZ is made for the S-parameter measurements as defined in Annex A of IEC/EN 61000-4-6 edition 4.
- Technical Specifications:
- Frequency range:
- 10 kHz to 1000 MHz
- Nominal impedance:
- 50 Ohms
- Connector:
- N-type female
- Maximum input level
- 0.01 to 100 MHz:
- 100 W for 15 min
- 100 to 230 MHz:
- 100 W for 3 min
- 230 to 1000 MHz:
- 50 W for 3 min
- Correction factor 0.15 to 300 MHz:
- >= -4 dB/<= 5 dB (see graph)
- Clamp opening diameter:
- approx. 21 mm
- Maximum cable diameter:
- 20 mm
- Dimension (LxWxH):
- 647 mm x 130 mm x 105 mm
- Height of ferrite center above ground:
- 50 mm
- Distance to the reference point (first ferrite):
- approx. 22 mm
- Weight:
- approx. 7 kg