Fischer FCC-801-M2-M3-16A Front Side
Fischer FCC-801-M2-M3-16A Side
Fischer FCC-801-M2-M3-16A Connectors

Fischer FCC-801-M2-M3-16A

Rent CDN for 2/3 Supply Lines

In Stock & Ready To Ship
Orders by 3pm ship today

The Fischer FCC-801-M2-M3-16A is a two- or three-line coupling decoupling network for IEC 61000-4-6 RF immunity testing, supporting 16 amps, 250 volts, and frequencies from 150 kHz to 230 MHz.

Features

  • Frequency 150 kHz - 230 MHz
  • Max 250 Volts 16 Amps AC/DC
  • Designed for 2 or 3 Supply Lines
  • Includes Calibration Adapters
Up to 4 Free Days of Transit Time
Two transit days each way are excluded from rental period
  • FCC-801-M2/M3-16A
  • 2 Golden Three Pin Adapters
  • 2 pcs 150/50 ohm Impedance
  • Printed Manual
  • Calibration Certificate
  • EUT Cable IN
  • EUT Cable Out
  • 50 ohm Terminator
  • Shipping Case
  • Can't find something? Contact Us

Description

The Fischer FCC-801-M2-M3-16A is a coupling decoupling network designed for conducted RF immunity testing of equipment connected to two- or three-line power systems. Operating from 150 kHz to 230 MHz, it injects controlled common mode disturbances in accordance with IEC 61000-4-6. The network supports operating voltages up to 250 volts and continuous currents up to 16 amps. A 50-ohm BNC connector provides the RF input, while safety sockets connect the equipment under test and auxiliary equipment. Its standardized 150-ohm common mode impedance and 10 dB voltage division factor support repeatable test levels, dependable decoupling, straightforward verification, and accurate compliance measurements.

Technical Specifications:
Application:
IEC 61000-4-6 Conducted Immunity
Frequency Range:
150 kHz to 230MHz
Supply Lines:
3 Power Conductors
Impedance:
150 kHz - 26MHz: 150 Ohm +/- 20 Ohm
26MHz - 80MHz: 150 Ohm + 60 Ohm -45 Ohm
80MHz - 230MHz: 150 Ohm +/- 60 Ohm
Standard
Published IEC 61000-4-6, Clause 6.2 Table 3 & Annex B Table B.1
Maximum Current:
16A
M2 Section Maximum Voltage:
DC - 60Hz 250V
M3 Section Maximum Voltage:
50Hz - 60Hz 250V
EUT/Mains Connector:
Multi-Contact Safety Socket 4mm
RF Disturbance Port:
50 Ohm BNC
Max Test Volts:
<30V
Fischer FCC-801-M2-M3-16A Datasheet
Download
Fischer FCC-801-M2-M3-16A User Manual
Download

Video Library

Related Videos

Rent the Fischer FCC-801-M2-M3-16A

Weekly, two-week and monthly rentals available.

Questions?630-887-0329