EM Test TVT 1-250-32
Related Equipment
Description
Voltage dips and short interruptions simulate a reduction of the power supply voltage for a certain period of time. Four different test levels are required as per IEC/EN 61000-4-11. The TVT 1-250-xx series of tapped transformer are used to simulate power supply failures such as voltage dips and interruptions providing taps with 40 %, 70 %, 80 % and a direct output with 100 % of the nominal mains supply voltage. Depending on the model the taps are selected manually or by means of the 0-10 V DC signal provided by the test generator.
- Technical Data:
- Input voltage UIN:
- max. 250 V (50 Hz/60 Hz)
- Output voltage UVAR:
- 100 % Vin, 80 % Vin, 70 % Vin and 40 % Vin
- Accuracy:
- 5 +/-%
- Voltage change with load:
- 100% output, 0 to 16 A less than 5%
80% output, 0 to 20 A less than 5%
70% output, 0 to 23 A less than 5%
40% output, 0 to 40 A less than 5% - Output current capability at 230 V input voltage:
- 16 Arms at 100% output
20 Arms at 80% output, for 5 s
23 Arms at 70% output, for 3 s
40 Arms at 40% output, for 3 s - Voltage selection:
- Touch screen user interface or WIN 3000 remote control software
- EUT power ON/OFF function:
- Front panel switch with power ON indicator circuit breaker operated by NSG 3000 user interface
- Fuse:
- 16 A, slow-blow
- Connectors:
- Harting type HAN3A, compatible with NSG 3000 safe power linking concept
- Possible extensions:
- Connector for optional remote EUT power-off switch, interlock connection to door switch
- Instrument supply:
- Selectable 100 to 110 VAC, 220 to 250 VAC, 15 VA, 47 to 63 Hz
- Dimensions:
- L x W x H 460 x 200 x 160 mm (18.1 x 7.9 x 6.3 Inches)
- Weight:
- 15 kg (33 lbs) approx.
- Cable:
- Power network receptacle to INA 6502 ~2 m (~79 Inches) (uses input cable delivered with NSG 3000 mainframe), INA 6502 to NSG 3000 EUT power input ~2 m (~79 Inches)
- Control cable:
- 2 m (79 Inches), 25 way sub D, twisted pair,shielded (included in delivery)