MegaPulse Defib D5-PF - Front
MegaPulse Defib D5-PF - Side 2
MegaPulse Defib D5-PF - Side
MegaPulse Defib D5-PF - Back

Compliance West MegaPulse D5-PF

Rent Defibrillation Tester
  • IEC 60601-2-25 and -2-27:2011 Generator
  • Integrated Energy Measurement
  • Output Voltage & Polarity From Front Panel
  • Y1-Y2 Measuring Circuit Integrated
  • Current ISO 17025 Accredited Calibration
Available Options

Included Accessories

  • Pulse Generator 
  • Printed Manual
  • Calibration Documentation
  • (3) HV Test Leads (Red)
  • (2) HV Test Leads (Black)
  • Software CD
  • USB Box
  • DB-9 Cable (Male to Male)
  • Case
Can't find something? Contact Us
Got Questions? Speak to an Expert +1 630-887-0329
Two FREE days transit with every rental

Description 

Surge tester built to the requirements of IEC 60601-1:2005 (Issue 3) Figure 10 (IEC 60601-1 Figure 50 in older versions). Conducts IEC 60601-1:2005 Figure 11; IEC 60601-2-34 Issue 2(Optional); IEC60601- 2-34 Issue 3(Included); IEC 60601-2-49; EC-13; EC-53 and the new requirement of IEC 60601-2- 25:2011 and IEC 60601-2-27:2011 including Correction 1. Tests of protection against the effects of defibrillation (Common mode, Differential mode), Energy reduction test. All these tests use the same 5000V source with difference in output wave shaping. The various wave shaping output circuits required for each test are included in the D5-PF.

The more remarkable changes of the D5-PF version are: output voltage selected digitally, direct joules readout, increased accuracy, faster time between pulses "20 seconds", increased signal generator voltage, pass/fail reference is integrated, dry cap is used and USB Testminder is standard.

Specifications
Charge Voltage:
0 - 5100 V tolerance +/-1%
Main Capacitance:
32 uF +/-5%
Inductances:
500uH and 25 mH +/-5%
Main Resistances:
100 Ohm, 50 Ohm and 400 Ohm +/- 1% non inductive
Voltage Control:
Manually, by turning a knob in the front panel or via USB
Polarity Control:
Positive and Negative. Alternating control available with USB
Voltage Display:
4 Digit LED Display
Voltage Meter resolution:
5V
Voltage Meter Accuracy:
1% +/-
Duty Cycle:
20s between pulses, +/-1% 100 resistor bank, up to 70 pulses
12s between pulses, +1% <-5% 100 resistor bank, continuous dut
Line Voltage:
120V AC, 50/60Hz *optional different line voltages available
Energy Measurement
Works only when:
25mH inductor selected (in accordance with the energy reduction test)
Display Resolution:
1 Joule
Repeatability:
1 Joule +/- between pulses