Compliance West MegaPulse D5-PF
Included Accessories
Can't find something? Contact UsRelated Equipment
Description
Surge tester built to the requirements of IEC 60601-1:2005 (Issue 3) Figure 10 (IEC 60601-1 Figure 50 in older versions). Conducts IEC 60601-1:2005 Figure 11; IEC 60601-2-34 Issue 2(Optional); IEC60601- 2-34 Issue 3(Included); IEC 60601-2-49; EC-13; EC-53 and the new requirement of IEC 60601-2- 25:2011 and IEC 60601-2-27:2011 including Correction 1. Tests of protection against the effects of defibrillation (Common mode, Differential mode), Energy reduction test. All these tests use the same 5000V source with difference in output wave shaping. The various wave shaping output circuits required for each test are included in the D5-PF.
The more remarkable changes of the D5-PF version are: output voltage selected digitally, direct joules readout, increased accuracy, faster time between pulses "20 seconds", increased signal generator voltage, pass/fail reference is integrated, dry cap is used and USB Testminder is standard.
- Specifications
- Charge Voltage:
- 0 - 5100 V tolerance +/-1%
- Main Capacitance:
- 32 uF +/-5%
- Inductances:
- 500uH and 25 mH +/-5%
- Main Resistances:
- 100 Ohm, 50 Ohm and 400 Ohm +/- 1% non inductive
- Voltage Control:
- Manually, by turning a knob in the front panel or via USB
- Polarity Control:
- Positive and Negative. Alternating control available with USB
- Voltage Display:
- 4 Digit LED Display
- Voltage Meter resolution:
- 5V
- Voltage Meter Accuracy:
- 1% +/-
- Duty Cycle:
- 20s between pulses, +/-1% 100 resistor bank, up to 70 pulses
12s between pulses, +1% <-5% 100 resistor bank, continuous dut - Line Voltage:
- 120V AC, 50/60Hz *optional different line voltages available
- Energy Measurement
- Works only when:
- 25mH inductor selected (in accordance with the energy reduction test)
- Display Resolution:
- 1 Joule
- Repeatability:
- 1 Joule +/- between pulses