Compliance West MegaPulse D8-PF - Front Side
Compliance West MegaPulse D8-PF - Front
Compliance West MegaPulse D8-PF - Back
Compliance West MegaPulse D8-PF - Front Top
Compliance West MegaPulse D8-PF - Front Side Above

Compliance West MegaPulse D8-PF

Rent Defibrillation Proof Tester
  • Compliant to IEC 80601-1-2-26
  • Newest Model of D5-PF Series Testers
  • Direct Energy Readout for Energy Reduction Test
  • Criteria Selected From Front Display
  • Current ISO 17025 Accredited Calibration
Available Options

Included Accessories

  • Pulse Generator 
  • Printed Manual
  • Calibration Documentation
  • (3) HV Test Leads (Red)
  • (2) HV Test Leads (Black)
  • Software CD
  • USB Box
  • DB-9 Cable (Male to Male)
  • Case
Can't find something? Contact Us
Got Questions? Speak to an Expert +1 630-887-0329
Two FREE days transit with every rental

Description 

Surge tester built to the requirements of IEC 80601-2-26:2020, includes all the functions of a standard D5-PF, IEC 60601-1:2005 (Issue 3) Figure 10 (IEC 60601-1 Figure 50 in older versions). Conducts IEC 60601-1:2012 Figures 9, 10 and 11; IEC 60601-2-34 Issue 2(Optional); IEC60601-2-34 Issue 3(Included); IEC 60601-2-49; EC-13; EC-53, IEC 60601-2-25:2011 and IEC 60601-2-27:2011 including Correction 1. Tests of protection against the effects of defibrillation (Common mode, Differential mode), Energy reduction test. All these tests use the same 5000V source with difference in output wave shaping. The various wave shaping output circuits required for each test are included in the D8-PF.

Specifications
Testing Circuitry
Charge Voltage:
5100V +/-1%.
Main Capacitance:
32uF +/-5%.
Inductances:
500uH and 25mH +/-5%
Main Resistance (R1):
100Ohms +/-1% non-inductive.
Output Impedance:
50Ohms and 400Ohms +/-1% non-inductive. Optional 10Ohms available.
Voltage Control:
Manually, by turning a knob in the front panel or via USB
Polarity Control:
Positive and Negative. Alternating control available with USB
Voltage Display:
4 Digit LED Display.
Voltage Meter resolution:
5V.
Voltage Meter Accuracy:
1% +/-
Duty Cycle:
20s between pulses, +/-1% 100 Ohm resistor bank, up to 70 pulses.
12s between pulses, +1% <-5% 100Ohm resistor bank, continuous duty
Energy Measurement
Works only when:
25mH inductor selected (in accordance with the energy reduction test).
Display Resolution:
1 Joule
Repeatability:
2 Joules +/- between pulses.
Reading:
From 3000V to 5100V only
Electrical
Line Voltage:
120V AC, 50/60Hz.
Safety
Safety Interlock
2-pin terminal block on rear panel; shorted for operation