Overview

This system all by EMC Partner consists of the MIG-OS-OS1 and DOW3000 pulse generators with the CDNs for damped oscillatory waves testing to all 5 associated frequencies of IEC 610000-4-18.  An ISO 17025 accredited calibration comes standard with the rental. This products easy to use software allows for quick and easy testing and changing of parameters. This test solution is ideal for testing to a variety of damped oscillatory testing including ANSI C37.90, IEC 61000-4-18, IEC 60255-22-1, and many others.

Specifications

Common Standards

EN/IEC 61000-4-18

External CDN:

Single or 3 phase coupler/decoupler with 4 I/O line paths

  AC Voltage and Current              

Up to 480VAC, 25A max AC

  DC Voltage and Current

115 VDC max. L to L, 60V L to N, 25A max.

1 MHz, 100kHz IEC Damped Oscillatory:

 

  Standard:

IEC 61000-4-18, IEC 60255-22-1


  Voltage Peak open circuit

250V – 10% to 3kV + 10%

  Voltage Risetime

75ns Rise Time ±20%

  Oscillation Frequencies

1MHz ± 10%, 100kHz ± 10%

  Output Impedance

200 Ω ± 20%

  Decay

50% of peak value between 3rd and 6th period

  Burst Duration

1  up to 99 s ± 10%

  Repetition rate maximum

400/s for 1MHz, 40/s for 100kHz

1 MHz ANSI / IEEE Damped Oscillatory:

 

  Standard:

ANSI / IEEE C37.90.1

  Voltage Peak open circuit

250V – 10% to 3kV + 10%

  Voltage Risetime

50 – 100 ns Rise Time

  Oscillation Frequencies

1MHz – 1.5MHz

  Output Impedance

150 Ω ± 20%

  Decay

50% of peak value after 6 µs

  Burst Duration

1  up to 99 s ± 10%

  Repetition rate maximum

400/s for 1MHz, 40/s for 100kHz

Weight – MIG-S-OS1

28kg, 62 pounds

Weight – CDN 2000-6-25

26kg, 57 pounds

Dimensions – MIG-OS-OS1

45x57x38cm, 17x22x15 inches

Dimensions - CDN 2000-6-25

45x50x19cm, 17x20x7.5 inches

3 MHz, 10 MHz, 30 MHz IEC Damped Oscillatory:

 

  Standard:

IEC 61000-4-18

  Voltage OC direct out

0.4 kV – 4.4 kV

 Voltage OC CDN out

0.4 kV – 4.0 kV

Voltage calibrated

0.5 kV – 4 kV ± 10 %


 Voltage waveform decay

Pk5 > ½ · Pk1, Pk10 < ½ · Pk1

  Output impedance

50 Ω

  Repetition rate maximum

400/s for 1MHz, 40/s for 100kHz

 Voltage rise time

5 ns ± 30 %

  Pulse repetition

max. 6666 / s

  Burst duration

1 ms – 20 s

Current waveform decay

Pk5 > ¼ · Pk1, Pk10 < ¼ · Pk1

 Burst repetition

10 ms – 200 s

  Current SC direct & CDN

8 A – 88 A @ all frequencies

  Current SC calibrated

10 A – 80 A ± 20 %

  Current rise time

< 330 ns @ 3 MHz < 100 ns @ 10 MHz < 33 ns @ 30 MHz

 Polarity

positive, negative, alternating

  Synchronization

0 – 359° ± 10°, 1° step

  Ramp

test voltage

 Optional accessories

CN-EFT1000, VERI01 OSI, VERI1K EFT, VERI50-EFT

Standards

Widely Used Standards

This list is a more comprehensive list of all the standards covered with this automotive load dump generator. This is by no means all the standards covered, but the major ones from the largest manufacturers and most widely used standards.

 

  • IEC 60255-4
  • IEC 6055-22-1
  • IEC 61000-4-18
  • IEC 60834-1
  • VDE 435 Teil 303