Description
Overview
The single phase motor variacs of the NX 1-260-XX series allow to steplessly adjust the EUT supply voltage from 0V to 260 V (280 V) when supplied by a 230 V mains supply voltage. By means of the variac NX 1-260-xx series tests as required in the standard IEC / EN 61000-4-11 (voltage dips and short interruptions as well as the optional voltage variation) can be performed adjusting the voltage to any voltage level. Apart from this the motor variac is used for power-frequency magnetic field tests as per IEC / EN 61000-4-8.
Data Sheet User ManualSpecifications
Input Voltage | Max 250 V |
Frequency |
50 Hz/60 Hz |
Output voltage |
0 - 130 V @115 V, 0 - 260 V @230 V, 0 - 275 V @250 V |
Output current |
Max. 16 A (variac NX1-260-16) |
Output power |
0 to 4,160 VA |
Technical Data |
|
Output PF1 |
Nominal mains supply voltage |
Output PF2 |
0 V - maximum output voltage (depending on the model and supplied mains supply voltage) |
Output connectors |
4 mm safety lab connectors PF1, PF2, Neutral and PE |
Dimensions |
19" / 6 HU 266 mm x 485 mm x 400 mm (HxLxW) |
Weight |
approx. 27 kg |
Supply voltage |
115/230 V |
Fuses |
20 A (PF1), 16 A (PF2) |