Schwarzbeck NNLK 8129
Related Equipment
Description
A V-LISN from the series of the NNLK line is required to measure conducted interference voltages according to VDE-, CISPR- and related standards. It provides the device under test with a mains connection that has a standardized impedance of 50 Ω || 50 µH and reduces interference voltages coming from mains. The interference voltage of the device under test is decoupled via a 7 kHz high pass filter to the EMI measurement receiver.
A path selector switch allows measuring both paths, one after another. The paths that are not being measured are internally terminated with 50 Ω
For measurements conforming to standards the references in the standard CISPR 16-2-1 have to be followed. Amongst others it is described that the AMN has to be grounded with a short and wide ground strap to the wall of the shielded chamber (or provisionally to a sheet metal board at the wall).
- Specifications:
- Frequency range:
- 150 kHz - 30 MHz
- AMN impedance:
- 50 uH || 50 Ohms
- Maximum continuous current:
- 4x 200 A
- Maximum voltage:
- 250 VAC 50/60 Hz, 400 VDC
- Maximum voltage when using option 400/700 V:
- 400 VAC 50/60 Hz, 800 VDC
- Standard:
- CISPR 16-1-2
- Connector for supply:
- Wing terminals 12 mm, removable
- Connector for EuT:
- Wing terminals 12 mm, removable
- Weight:
- Roughly 30 kg
- Dimensions housing incl. wing terminals (WxHxD):
- 448 mm x 383 mm x 605 mm
- Connector to EMI receiver:
- BNC socket, 50 Ohms