- MANUFACTURER
- EMC Partner
- MODEL
- IMU-MGE F5-S6-R6
- TEST FUNCTIONS
- EFT burst and combination wave surge and ring wave
- TEST STANDARDS
- IEC 61000-4-4 and IEC 61000-4-5 and IEC 61000-4-12
- EFT OPEN CIRCUIT VOLTAGE
- 0.2 to 5.3 kilovolts plus or minus 10 percent
- EFT VOLTAGE INTO 50 OHMS
- 0.1 to 2.65 kilovolts plus or minus 10 percent
- EFT OUTPUT IMPEDANCE
- 50 ohms
- EFT RISE TIME INTO 50 OHMS
- 5 nanoseconds plus or minus 1.5 nanoseconds
- EFT PULSE DURATION INTO 50 OHMS
- 50 nanoseconds plus or minus 15 nanoseconds
- EFT SPIKE FREQUENCY
- 1 kilohertz to 1 megahertz
- EFT BURST DURATION
- 0.01 to 30 milliseconds with continuous operation up to 8 kilohertz
- EFT BURST REPETITION PERIOD
- 1 to 1000 milliseconds
- SURGE OPEN CIRCUIT VOLTAGE
- 0.25 to 6.6 kilovolts plus or minus 10 percent
- SURGE SHORT CIRCUIT CURRENT
- 125 to 3300 amperes plus or minus 10 percent
- SURGE VOLTAGE WAVEFORM
- 1.2 microsecond front time plus or minus 30 percent and 50 microsecond duration plus or minus 20 percent
- SURGE CURRENT WAVEFORM
- 8 microsecond front time plus or minus 20 percent and 20 microsecond duration plus or minus 20 percent
- SURGE OUTPUT IMPEDANCE
- 2 ohms
- SURGE REPETITION
- Up to 1 pulse per second at 0.6 kilovolts and 1 pulse per 10 seconds at 6.6 kilovolts
- RING WAVE OPEN CIRCUIT VOLTAGE
- 0.25 to 6.6 kilovolts plus or minus 10 percent
- RING WAVE FREQUENCY
- 100 kilohertz plus or minus 10 percent
- RING WAVE VOLTAGE RISE TIME
- 0.5 microseconds plus or minus 30 percent
- RING WAVE OUTPUT IMPEDANCE
- 12 ohms or 30 ohms
- RING WAVE SHORT CIRCUIT CURRENT AT 12 OHMS
- 20.833 to 550 amperes plus or minus 10 percent
- RING WAVE SHORT CIRCUIT CURRENT AT 30 OHMS
- 8.3 to 220 amperes plus or minus 10 percent
- RING WAVE REPETITION
- Up to 1 pulse per second at 6.6 kilovolts
- PULSE POLARITY
- Positive and negative and alternating
- PHASE SYNCHRONIZATION
- 0 to 359 degrees in 1 degree steps
- INTEGRATED COUPLING DECOUPLING NETWORK
- Single phase rated for 16 amperes
- EUT AC VOLTAGE
- Up to 300 volts AC at 50 or 60 hertz
- EUT DC VOLTAGE
- Up to 90 volts DC as standard and up to 300 volts DC with optional EXT-IMU P or EXT-IMU D module
- EFT COUPLING
- Line and neutral and protective earth individually or in combinations and direct output
- SURGE COUPLING
- Line to neutral at 2 ohms and line or neutral to protective earth at 12 ohms and direct output
- RING WAVE COUPLING
- Line to neutral and line to protective earth and neutral to protective earth at 12 ohms or 30 ohms and direct output
- DISPLAY
- 7 inch color capacitive touchscreen
- INTERFACES
- Gigabit Ethernet and USB and RS485
- EUT MONITORING
- Supply voltage and current and frequency displayed on screen
- ENVIRONMENT MONITORING
- Built in temperature and humidity sensor
- OPERATING SYSTEM
- EPOS proprietary firmware
- SAFETY
- Front panel emergency stop button and safety circuit
- PRIMARY POWER
- 100 to 240 volts AC plus or minus 10 percent at 50 or 60 hertz
- POWER CONSUMPTION
- Less than 150 volt amperes in operation and less than 15 volt amperes in standby
- MAINFRAME WEIGHT
- 30 kilograms excluding installed modules
- SIZE WIDTH BY HEIGHT BY DEPTH
- 600 by 370 by 450 millimeters
- OPERATING TEMPERATURE
- 10 to 35 degrees Celsius
- OPERATING HUMIDITY
- Less than 80 percent noncondensing
- AIR PRESSURE
- 86 to 106 kilopascals