Teseq NSG 3060A Front Side
Teseq NSG 3060A Front
Teseq NSG 3060A Back
Teseq NSG 3060A Front Top
Teseq NSG 3060A Front Side Above

Teseq NSG 3060A

Rent 6 kV Transient Test System
  • Configured for 6.6 kV Surge & 5.5 kV EFT/Burst
  • Built-in single phase CDN Max 300 V/16 A AC or DC
  • Model: NSG 3060A-IEC (IEC 61000-4-4/5)
  • Newest Generator in Teseq NSG Series
  • Current ISO 17025 Accredited Calibration

Included Accessories

  • Immunity System
  • USB drive with manual 
  • Power Cord
  • Printed User Manual
  • Calibration Documentation
  • SWL AD Dummy Connector 
  • SCT 112801 Dummy Connector
  • EUT IN Power Cable
  • SHV to SHV Cable
  • Ground Attachments 
  • ESA 1.6 Adapter Gray Box
  • Shipping Case 
Can't find something? Contact Us
Got Questions? Speak to an Expert +1 630-887-0329
Two FREE days transit with every rental

Description

Teseq’s new NSG 3060A conducted immunity generator takes the proven, user-friendly design of the highly successful NSG 3000 series to a new level. This innovative design uses modular architecture to provide a versatile system that can be configured for basic testing needs and expanded to meet the needs of sophisticated test laboratories.

Designed to fulfill requirements for CE mark and ANSI C62.41 testing, the NSG 3060A performs tests for Combination wave surge, Ring wave and Electrical Fast Transient (EFT) pulses as well as Power Quality Testing (PQT). Extensive expansion capabilities enable the system to be configured for a much broader range of applications.

Using state of the art components, the self-contained modules set new standards with respect to switching and phase accuracy and exceed the existing standards’ requirements. With its powerful processors, the NSG 3060A can completely fulfill the unique coupling requirements specified by ANSI C62.41. This standard requires that the pulse amplitude be adjusted for the phase position of the pulse on the AC mains, and for the amplitude of the mains voltage.

A 7” touch panel display with superb contrast and color is the most striking feature of the new NSG 3060A. For fast and efficient data entry, input devices include an integrated keyboard and a thumbwheel with additional keys for sensitivity adjustment.

The user-friendly graphic display speeds test setup. Each parameter’s value is highly visible, and all settings can be quickly selected and modified with the generously sized touch input buttons. A stylus is not necessary, and ramp functions are programmed quickly and easily. Multi-step test procedures can be created and their sequence or parameter values changed easily.

The users can make manual parameter changes using the thumbwheel while a test is under way, providing an effective and fast method for identifying critical threshold values. The Test Assistance (TA) function allows users to initiate standardized test with just a few “clicks” to achieve quick, reliable results in a development environment.

The NSG 3060A has an Ethernet port for external PC control. The Windows-based control software simplifies test programming and allows compilation of complex test sequences with diverse pulse types. Test reports can be generated during the test operation, allowing the operator to enter observations as the test progresses and increasing the efficiency of long-term test

Specifications
Combination wave pulse:
Pulse voltage (open circuit):
200 V +- /- to 6.6 kV (in 1 V steps)
Pulse current (short circuit):
100 A +/- to 3.3 kA
Impedance:
2/12 Ohms
Polarity:
positive / negative / alternate
Pulse repetition:
10 s* up to 9,999 s (in 1 s steps)
Test duration:
1 to 99,999 pulses, continuous
Phase synchronization:
asynchronous, synchronous 0 to 359 degrees (in 1 degree steps)
Coupling:
ANSI / IEC / external
Ring wave:
Pulse voltage (open circuit):
200 V +/- to 6.6 kV (in 1 V steps)
Pulse current (short circuit):
16.6 +/- to +/- 550 A , +/- 10%
+/- 6.6 to +/- 220 A, +/- 10%
+/-1 to +/- 33 A, +/-10%
Impedance:
12/30 ?
Polarity:
positive / negative / alternate
Pulse repetition:
10 s* up to 9,999 s (in 1 s steps)
Test duration:
1 to 99,999 pulses, continuous
Phase synchronization:
asynchronous, synchronous 0 to 359 degrees (in 1 degree steps)
Coupling:
ANSI / IEC / external
Burst (EFT):
200 V +/- to 4.8 kV (in 1 V steps) - open circuit
+/- 100 V to 2.4 kV (50 ? matching system)
Pulse amplitude:
100 Hz to 1000 kHz
Burst frequency:
positive / negative / alternate
Polarity:
10 ms to 9'999 ms
Burst time:
0.01 ms to 9'999 ms, single pulse
Test duration:
1 to 99,999 pulses, continuous
Phase synchonization:
asynchronous, synchronous 0 to 359 degrees (in 1 degree steps)
Coupling:
internal / external
Internal coupling network:
EUT supply:
1-phase
EUT VAC:
Up to 300 Vrms *, 50 / 60 Hz (phase - neutral)
EUT VDC:
Up to 300 VDC
EUT current
1 x 16 Arms continuous (over heat protected)
Surge
Standard coupling as per IEC 61000-4-5
Coupling mode
Line to Line
Line(s) to ground
Mains decoupling
1.5 mH 0% + 35%
Decoupling attenuation:
Residual pulse voltage on EUT power supply inputs 15 % max.
Residual voltage on non-pulsed EUT power supply inputs 15 % max
EFT (Burst)
Standard coupling all lines to ref ground (GND) IEC / EN 61000-4-4