Description
Overview
Surge tester built to the requirements of IEC 80601-2-26:2020, includes all the functions of a standard D5-PF, IEC 60601-1:2005 (Issue 3) Figure 10 (IEC 60601-1 Figure 50 in older versions). Conducts IEC 60601-1:2012 Figures 9, 10 and 11; IEC 60601-2-34 Issue 2(Optional); IEC60601-2-34 Issue 3(Included); IEC 60601-2-49; EC-13; EC-53, IEC 60601-2-25:2011 and IEC 60601-2-27:2011 including Correction 1. Tests of protection against the effects of defibrillation (Common mode, Differential mode), Energy reduction test. All these tests use the same 5000V source with difference in output wave shaping. The various wave shaping output circuits required for each test are included in the D8-PF.
Data SheetUser ManualSpecifications
Testing Circuitry | |
Charge Voltage: | 0-5100 V ± 1% |
Main Capacitance: | 32µF ± 5% |
Inductances: | 500uH and 25mH ±5%. |
Main Resistance (R1): | 100Ω ±1% non-inductive. |
Output Impedance: | 50Ω and 400Ω ±1% non-inductive. Optional 10Ω available |
Voltage Control: | Manually, by turning a knob in the front panel or via USB |
Polarity Control: | Positive and Negative. Alternating control available with USB. |
Voltage Display: | 4 Digit LED Display. |
Voltage Meter resolution: | 5V |
Voltage Meter Accuracy: | ±1% |
Duty Cycle: |
20s between pulses, ±1% 100Ohm resistor bank, up to 70 pulses. 12s between pulses, +1% <-5% 100Ohm resistor bank, continuous duty |