Description
Overview
The Model CI00402 is a fully self-contained state of the art system designed to test RF Conducted Immunity. The system contains a signal generator, spectrum analyzer, 100W nominal AR amplifier, a 10 kHz to 400MHz directional coupler and emcware® control software. Everything is contained in a single housing, which eliminates setup issues. This system provides the versatility needed for every test laboratory and equipment manufacturer. The RF amplifier, signal generator and spectrum analyzer can be used independently of the system. If special needs arise or standards were to change a larger amplifier can be connected to the system. The use of spectrum analyzer and signal generator may also be controlled by software.
Option 1 adds data acquisition capability. Option 2 provides a laptop PC. Option 3 is a lower-cost solution providing all the functions and features of the full CI00402 base model without the cost of the internal amplifier.
The export classification for this equipment is EAR99. These commodities, technology or software are controlled for export in accordance with the U.S. Export Administration Regulations. Diversion contrary to U.S. law is prohibited.
Data SheetUser ManualSpecifications
Common Standards |
EN/IEC 61000-4-6, MIL-STD-461 CS114, ISO 11452-4, & DO 160 sec 20 |
Signal Generator Specifications |
|
Frequency range resolution |
9 kHz to 1.5 GHz 0.01Hz |
Power range resolution |
-110 to +13 dBm 0.01dB |
Modulation |
AM, FM, Phase, Int Pulse, Ext Pulse |
Spectrum Analyzer Specifications |
|
Frequency range resolution |
9 kHz to 1.5 GHz 1 Hz |
RF Power CW (max) Atten = 30 dB |
20 dBm |
Resolution BW |
10 Hz to 1 MHz |
Video BW |
1 Hz to 3 MHz |
Amplitude Measurement Range |
-110 dBm to +20 dBm in 1 dB steps |
Preamplifier Gain |
20 dB (nom) |
Sweep Time, span> 100 Hz |
10 msec to 1500 sec |
RF Amplifier Specifications |
|
Frequency range |
10 kHz to 400 MHz |
Power rating |
100 Watts Minimum |
1dB compression |
75 Watts Minimum |
Harmonic Distortion |
-20dBc at 75 Watts |
Mismatch tolerance |
100% of rated power without fold back. Will operate without damage or oscillation with any magnitude of source and load impedance. |
Gain |
50 dB minimum |
Control Interface |
|
USB B |
Connection to the computer |
General |
|
Power |
115/230 VAC, 50/60 Hz, single phase 16 A |
Breaker |
2 pole, 20 A |
Dimensions |
128.9 x 56.1 x 91.4 cm (52.5 x 22.1 x 36 in.) |
Weight |
72.6 kg (160 lb.) |
Standards
Widely Used Standards
This list is a more comprehensive list of all the standards covered with this conducted RF Generator. For assured compliance to a particular standard, confirm prior to rental. This conducted immunity generator has the ability to either cover all, or part of the standards below.
- MIL-STD-461 (CS114)
- DO160 (Sec 20 BCI Test)
- IEC/EN 60601-1-2
- IEC/EN 50130-4
- IEC/EN 61326
- IEC/EN 61000-6-1
- IEC/EN 61000-6-2
- IEC/EN 61000-4-6
- CISPR 24/EN 55024
- ISO 11452-4
- EMC-CS-2009
- GM GMW3097
- BMW GS95002
- Chrysler DC-11224
- Renault 36-00-808