Teseq MD 102A
Related Equipment
Description
The MD 102a, successor to the MD 103 , is a electrostatic discharge calibration target used for verifying waveforms of ESD simulators/guns. All the elements incorporated within the measurement chain must be matched to one another before highly accurate measurements with minimal uncertainties can be contemplated. This test accessory offers distortion-free, flat frequency characteristic up to at least 6 GHz.
Faster oscilloscopes enable more complete measurements and analyses of the fast ESD pulses but, of course, the rest of the components in the measurement chain must be up to the necessary standard and exhibit the appropriate frequency response. Therefore the MD 102 set offers a complete measuring path from the target to the attenuator up to the high quality coax cable. The whole path has to be left together and should not be disassembled into parts, as the whole coupling path is calibrated in one piece. The MD102 replaces the MD101 & MD103 ESD targets.
- Specifications:
- Variation of insertion loss
- <0.2 dB up to 1 GHz
- .S21 parameter
- <0.3 dB up to 4 GHz
- <0.5 dB up tp 6 GHz