AE Techron T1000

Rent Magnetic Field Field Transformer
In Stock - Order by 3 p.m. ships today
  • DO 160 Section 19 & Boeing D6-16050-5 Section 7.2
  • Up to 100 A RMS Secondary Current
  • Output via Multi Contact 125A plugs
  • Optimized for use with AE Techron 7220 Series Amplifiers
Got Questions? Speak to an Expert +1 630-887-0329
Two FREE days transit with every rental

Description

The T1000 Magnetic-Field Susceptibility Transformer was designed to exceed the AF Magnetic Field susceptibility test requirements of DO 160 Section 19 and Boeing D6-16050 Section 7.2. It may also be used as a 10-to-1 step-down transformer for output current of up to 100ARMS.

The T1000 transformer provides up to 300A-m and is able to generate up to 100ARMS. The turns ratio provides a ten-to-one step down to the secondary winding.

The T1000 provides convenient input connectors via binding posts. Standard 0.75-inch spacing of binding posts allows use of standard plugs. Output is via Multi Contact 125A plugs.

A 12-amp circuit breaker protects the unit against over-current conditions. A rugged, impact-resistant case and robust design protects the transformer from accidental damage.

The T1000 transformer is for use by experienced staff.

Primary:
200 VP maximum
Secondary:
20.0 VP maximum output
Dielectric Test:
1200 VDC
Turns Ratio::
Ten-to-one step down
Protection:
Input power-protected via circuit breaker
Input Terminals:
4-way binding posts
Output Terminals:
Multi Contact 125A plugs