Amplifier Research FL8009/kit
Related Equipment
Description
The FL8000 series is an advanced line of laser-powered Electric Field Probes, and associated Kits designed specifically for EMC radiated immunity test applications. Each Field Probe is an electrically small, separable axis sensor capable of measuring both continuous wave and pulsed electric fields. There are a total of five Field Probes available, each with a unique bandwidth, sensitivity, and dynamic range.
The filed probe kits contain everything needed to operate its associated Field Probe. Remote control and laser power for the Field Probe is provided by the FI8000 Laser Probe Interface. A set of 20- meter fiber-optic extension cables and fiber-optic mating adapter are also provided with the kit. The use of a laser to deliver power-over-fiber (PoF) allows for continuous operation without recharging or replacing batteries. Additionally, this provides excellent RF isolation for the control system.
An ortho-angle mount and adapter foot are provided for ease of mounting in common orientations. When rotated about its ortho-angle mount, the probes provide excellent isotropic response. Attention was given to the mounting geometry and remote command set for the FL8000 series kits to allow seamless substitution with the legacy field probe kits.
Correction factors from an accredited ISO 17025 calibration lab are provided with each Field Probe. These factors can be loaded into a model FM7004A Field Monitor (sold separately) to automatically correct the probe readings at user-specified frequencies. When correction factors are applied, the true accuracy of the Field Probe can be realized.
The export classification for this equipment is EAR99. This equipment is controlled for export in accordance with the U.S. Export Administration Regulations. Diversion contrary to U.S. law is prohibited
- Specifications:
- Frequency Range:
- �20 MHz - 9.3 GHz
- Axis Type
- Separable X-Y-Z Axis
- Measurement Type
- CW, AM & Pulse
- Field Strength Range (Single Range)
- 0.5 - 800 V/m
- Dynamic Range
- >64 dB
- Analog Rise Time (10 � 90% Typical)
- 300 ns
- Isotropic Deviation (Measured at Ortho Angle)
- 0.5 dB +/- @ 100 MHz
- Resolution
- <0.1 dB
- CW Damage Level
- 1000 V/m
- Pulse Damage Level
- 5 kV/m (> 0.1% Duty)
- Linearity Error
- 0.5 dB +/- or 0.3 V/m +/- (Whichever is greater) (+/- 2.0 dB 20 MHz - 80 MHz)
- Temperature Stability (Over Operating Temperature Range
- 0.1 dB +/- (Detection Circuit)
- 0.5 dB +/-(Complete System)
- Weight
- 150 g (5.3 oz)
- Dimensions (W x H x D)
- 42.3 x 52.4 x 52.4 mm (1.66 x 2.06 x 2.06 in) 29.2 mm (1.15 in) Spherical housing diameter 16.5 mm (0.65 in) Sensor radome height per axis
- Data Response
- Internal Sampling Rate
- 2 Msps
- Sampling Window
- 4 ms
- Time Between Sampling Windows
- 30 ms
- Minimum Pulse Width
- 1 us
- Amplitude Accuracy (field aligned with sensor axis)
- Without Correction Factors Applied
- 1.0 dB +/- @ 100 MHz
- With Correction Factors Applied � Typical expanded measurement uncertainty � 95% Confidence Interval (Calibration Lab Dependent)
- 1.1 dB +/- (20 MHz - 1 GHz) +/- 2.1 dB (1 GHz - 9.3 GHz)
- Environmental
- Operating Temperature
- 10 to +40 C
- Storage Temperature
- 0
- Operating Altitude
- Up to 2000 m
- Shock & Vibration
- Normal Truck Transport
- Fiber-Optics
- Connector Type
- E � 2000 Compact Duplex
- Bend Radius
- >30 mm
- Pigtail Length (With Verification Loop)
- 1 m
- Maximum Single F/O Extension Length
- 100 m Contact AR Applications Team for requirements greater than 100 m