Compliance West MegaPulse D8-PF
Included Accessories
Can't find something? Contact UsRelated Equipment
Description
Surge tester built to the requirements of IEC 80601-2-26:2020, includes all the functions of a standard D5-PF, IEC 60601-1:2005 (Issue 3) Figure 10 (IEC 60601-1 Figure 50 in older versions). Conducts IEC 60601-1:2012 Figures 9, 10 and 11; IEC 60601-2-34 Issue 2(Optional); IEC60601-2-34 Issue 3(Included); IEC 60601-2-49; EC-13; EC-53, IEC 60601-2-25:2011 and IEC 60601-2-27:2011 including Correction 1. Tests of protection against the effects of defibrillation (Common mode, Differential mode), Energy reduction test. All these tests use the same 5000V source with difference in output wave shaping. The various wave shaping output circuits required for each test are included in the D8-PF.
- Specifications
- Testing Circuitry
- Charge Voltage:
- 5100V +/-1%.
- Main Capacitance:
- 32uF +/-5%.
- Inductances:
- 500uH and 25mH +/-5%
- Main Resistance (R1):
- 100Ohms +/-1% non-inductive.
- Output Impedance:
- 50Ohms and 400Ohms +/-1% non-inductive. Optional 10Ohms available.
- Voltage Control:
- Manually, by turning a knob in the front panel or via USB
- Polarity Control:
- Positive and Negative. Alternating control available with USB
- Voltage Display:
- 4 Digit LED Display.
- Voltage Meter resolution:
- 5V.
- Voltage Meter Accuracy:
- 1% +/-
- Duty Cycle:
- 20s between pulses, +/-1% 100 Ohm resistor bank, up to 70 pulses.
12s between pulses, +1% <-5% 100Ohm resistor bank, continuous duty - Energy Measurement
- Works only when:
- 25mH inductor selected (in accordance with the energy reduction test).
- Display Resolution:
- 1 Joule
- Repeatability:
- 2 Joules +/- between pulses.
- Reading:
- From 3000V to 5100V only
- Electrical
- Line Voltage:
- 120V AC, 50/60Hz.
- Safety
- Safety Interlock
- 2-pin terminal block on rear panel; shorted for operation