Rent Close Proximity Magnetic Field System
  • 9 kHz to 50 MHz 100 Watt Power Amplifier
  • IEC 61000-4-39 & IEC 60601-1-2 Immunity Tests
  • Rental Antenna Test Sets Available
  • Includes Laptop with icd.control Software
  • Current ISO 17025 Accredited Calibration
Got Questions? Speak to an Expert +1 630-887-0329
Two FREE days transit with every rental

Description

The NSG 4070C-LFCP is a universal device for standard-compliant and development-accompanying EMC immunity tests in the application area of magnetic field testing in the close proximity according to IEC / EN 61000-4-39 and IEC / EN 60601-1-2. The NSG 4070C-LFCP integrates signal generator, directional coupler, power amplifier, power meter and EUT monitoring interfaces. The NSG 4070C-LFCP can be operated quickly, conveniently and easily via the front panel as a free-standing device. Test and measurement data can be conveniently transferred for documentation purposes via a USB stick. The NSG 4070C-LFCP can be remote controlled via LAN, electrical or optical RS232 as well as USB.

The wide frequency range of the signal generator and the power meter offer the possibility to connect external amplifiers and directional couplers. In combination with a remote control software, external power amplifiers and directional couplers, further applications can be covered, such as testing of conducted disturbances induced by high frequency fields according to IEC / EN 61000-4-6, automotive BCI tests e.g. according to ISO 11452-4 as well as high frequency electromagnetic fields according to IEC / EN 61000-4-3 or IEC / EN 61000-4-20. For EUT monitoring, the NSG 4070C-LFCP offers a variety of interfaces for flexibility in laboratory use.

In order to start with predefined parameter settings the optional test software icd.control is recommended. The software offers a large standard database and predefined drives for using external measuring devices. More complex systems including radiated tests can be controlled by using the software solution CIS (Compliance Immunity Software). It also allows for all coupling methods to be simply and easily connected and calibrated.

Specifications:
Generator
RF
Frequency Range
4 kHz to 1 GHz
Resolution
1 Hz
Reference Frequency
10 MHz
Aging
25 ppm
RF Level
Level range:
-60 dBm to +10 dBm
Resolution:
0.1 dB
Settling time:
10 ms
Amplitude modulation
Modulation depth:
0 to 100%
Modulation frequency range:
1 Hz to 50 kHz
Frequency resolution:
1 Hz
External modulation
Delay time:
< 1 us /180 degrees
Period:
min. 20 us
Pulse width:
min. 10 us
Power meter
Frequency range:
4 kHz to 1 GHz
Linear measurement range
channel 1:
-35 dBm to +27 dBm
channel 2,3:
-45 dBm to +20 dBm
Max. input/no damage
Channel 1-3:
+28 dBm
Noise level:
>5 dB below the measurement rang
Input return loss:
>20 dB (below 500 MHz), >17 dB (500 MHz to 1 GHz)
Connector:
BNC socket, 50 Ohm
Accuracy 10 to 30 C:
<0.5 dB, typ. <0.3 dB
Power Amplifier
Nominal output power:
100 W
Frequency range:
9 kHz - 50 MHz
Type:
single band, class A
Input/output impedance (nominal):
50 Ohm
Input return loss (minimum):
10 dB
Output return loss (nominal):
min. 9.5 dB
Output return loss without damage:
0 dB
Gain (minimum):
50 dB
Gain flatness (maximum):
+/- 3 dB