EMC Partner IMU4000 - Front Side
EMC Partner IMU4000 - Front
EMC Partner IMU4000 - Back
EMC Partner IMU4000 - Front Top
EMC Partner IMU4000 - Front Side Above

EMC Partner IMU4000

Rent 4 kV Surge & 4.4 kV EFT Generator

  • 4 kV Surge & 4.4 kV EFT/Burst Capabilities
  • CDN Max 280 V/16 A AC or 300 V/16 A DC
  • Capabilities for Voltage Dips up to 6 Amps
  • Replaced by IMU-MGS Series of Generators
  • Current ISO 17025 Accredited Calibration

Included Accessories

  • Power Cord
  • EUT Power In Cable with MC Connector
  • EUT Power out Cable
  • Red Ethernet Cable
  • (2) U shaped jumpers(attached on back)
  • EU to USA adapter plug
  • (2)CD with Software and Manuals
  • Software on Flash Drive
  • Printed Manual
  • Calibration Papers
  • Pelican Shipping Case
Can't find something? Contact Us
Got Questions? Speak to an Expert +1 630-887-0329
Two FREE days transit with every rental

Description

The tester IMU4000 simulates transients of different interference sources. Such as: indirect lightning in electronic systems, human body electrostatic discharges, switched inductance (Burst), power supply interruptions and variations and common mode disturbance.

The test system IMU4000 with accessories fulfils all requirements of the IEC basic standards 61000-4-2 Ed.2 (ESD); 61000-4-4 Ed.3 (EFT); 61000-4-5 Ed.3 (SURGE); impulse; 61000-4-11 Ed.2 (Interruption and Variations), and with accessories 61000-4-8 Ed.2 (Magnetic field 50/60Hz) and 61000-4-9 Ed.1 (Magnetic field SURGE), 61000-4-29 Ed.1 dips and interruption on d.c. and IEC 61000-4-34, DIPS and Interruption >16A per phase.

If not all transient test are needed, the IMU4000 tester is also available in various versions, with the possibility to upgrade the tester later to a full IMU4000 test system.

Mainframe Specifications:
IMU4000 Control features
User Interface
7 Inch Colour touch panel
Operating System
EMC PARTNER (EPOS)
Communication Interface
Ethernet
Accessory control interface
RS485
Atmospheric measurement
Temperature, humidity, pressure (with accessory)
BNC monitor ports
EUT plus IMPULSE voltage & current
Trigger mode
Auto, manual, external
Synchro source
EUT power, Impulse out, External
Synchro on / off
0 Degrees up to 360 Degrees
IMU4000 Coupling Decoupling Network
Maximum AC voltage
280V
Maximum DC voltage
300V
Maximum EUT current
16A
Frequency range
DC up to 60Hz
Power frequency synchro
16.7 up to 60Hz
Coupling EFT
L, N, PE, L+N, L+PE, N+PE, L+N+PE, direct
Coupling CWG
L-N, L-PE, N-PE, direct
Electrical Fast Transient (Burst)
EXT-TRA3000 F
EFT 4 kV
Voltage range
0.25 up to 4.4kV
Voltage amplitude into 50ohm
0.125 up to 2.0kV
Voltage amplitude into 1kohm
0.24 up to 4.0kV
Surge Impulses
EXT-TRA3000 S
CWG 4 kV
Voltage range
0.25 up to 4.1kV
Current range
0.125 up to 2.05kA
Source impedance
2ohm
Serial resistance common mode
10ohm
Pulse front time at open circuit
1.2us
Pulse duration at open circuit
50us
Pulse front time at short circuit
8us
Pulse duration at short circuit
20us
Pulse repetition
up to 60 pulses per minute
Pulse repetition at maximum voltage
6 pulses per minute
Polarity
Positive / Negative / Alternating
Programmable parameter ramps
voltage, synchronisation, polarity
Synchronisation on power line frequencies
Yes
Voltage Dips
EXT-TRA3000 D
AC Dips
Voltage range
0 up to 250V
Frequency range with variac
48 up to 60Hz
Nominal current
16A
Interruption time
50us up to 30s
Amplitude interrupt with internal variac
0 up to 100% max. 5A
Phase angle to turn on and off
0 Degrees up to 360 Degrees
Voltage variation with internal variac
0 up to 110% max. 5A
Interrupt less than 1 period
Input by phase angle
Interrupt more than 1 period
Input in milli seconds
Programmable parameter ramps
voltage, synchronisation, interrupt
Rise and fall time at 100ohm load
1 up to 5us
Option 3.2 Special Program
50ms Interrupt test to IEC 62052-11