Description
Overview
Rent the CWS 500 N1.4 by the most common of the CWS 500N series by EM Test designed to provide testing solutions to IEC 61000-4-6 with a 9kHz-1Ghz built in signal generator. We rent a variety of accessories and test equipment associated with this series of RF immunity including the CWS 500N2 and many testing accessories. All of our RF immunity generators come with ISO 17025 accredited calibration and technical support.
The CWS 500N 1.4 was created to test to EN/IEC 61000-4-6, IEC 60601-1-2, ISO 11452-4, and many others. This test system has a diverse range of applications for testing to the 1kHz 80% Am singal as well as a 2Hz 80% AM signal to test medical appliances. This unique single box solution provides software allowing for many reprogrammed standards to be run efficiently.
The EM Test CWS 500n1 is also equipped with a 1 GHz current monitor the can be used up to 1 GHz by means of an external amplifier. As an added bonus the CWS 500n1 provides a 1Hz PM signal with 50% duty cycle required for test safety equipment. A diverse set of accessories are available for rent with the CWS 500N1 meeting any testing needs.
Data Sheet User ManualSpecifications
Common Standards |
EN/IEC 61000-4-6, IEC 60601-1-2, ISO 11452-4 |
Test Level |
|
Output Voltage Level |
1V – 30V |
Output Power |
80W (nominal) |
Output Impedance |
50 Ω |
Harmonic Distortion |
Less than -20dBc at max. power |
Test Frequencies |
|
Sinus (CW) |
100kHz to 300MHz |
Frequency Bands |
100kHz to 9.999MHz 10MHz to 99.99Mhz 100Mhz to 300MHz |
Unmodulated Signal |
CW (continuous wave) |
Amplitude Modulation |
1kHz, 80%AM as per IEC 61000-4-6 2H, 80%AM as per IEC 60601-1-2 400Hz, 80% AM |
Pulse Modulation |
1 Hz, 50% duty cycle (EN 50130-4) |
Measurements |
|
Monitor |
Integrated RF power meter, measuring input for CDN and clamp calibration as well as current monitor for clamp applications |
Cal data F1-F5 |
5 internal memories to save calibration data |
Time Parameters |
|
Dwell Time for CW & AM |
td = 0.3s to 9,999s |
Dwell Time for PM |
td = 3s to 9,999s |
Pause Time |
tr = 0/0.3s to 9,999s |
RF Output |
Direct RF output, N Connector(on front panel) |
Test Routines |
|
User Test Routines |
Voltage sweep, Frequency sweep, Dwell time sweep |
Standard Test Routines |
IEC 610000-4-6 , level 1 to level 3,Automatic Level X and Y |
Signal Generator |
|
Output level |
-63.5dBm to 0dBm |
Frequency Range |
9kHz to 1GHz |
Output Impedance |
50 ohm |
Direct RF Output |
To control an external amplifier |
Weight |
17 kg, 37 pounds |
Dimensions |
44x48x14 cm,17x19x6 inches |
Standards
Widely Used Standards
This list is a more comprehensive list of all the standards covered with this conducted RF Generator. For assured compliance to a particular standard, confirm prior to rental. This conducted immunity generator has the ability to either cover all, or part of the standards below.
- BMW GS 95002 (2001)
- DaimlerChrysler DC-10614
- DaimlerChrysler DC-11224
- EN 300329
- EN 300340
- EN 300342-1
- EN 300386 V1.3.2
- EN 301489-1
- EN 301489-17
- EN 301489-24
- EN 301489-7
- EN 55024
- EN 61000-6-1
- EN 61000-6-2
- Fiat 9.90110
- Ford ES-XW7T-1A278-AB
- Ford ES-XW7T-1A278-AC
- GMW 3097 (2006)
- IEC 60601-1-2
- IEC 61000-4-6
- IEC 61326
- IEC 61850-3
- ISO 11452-4
- Mercedes MBN 10284-2
All Accessories
Rent a variety of Conducted RF Immunity accessories with test solution.
Take advantage of A2LA ISO 17025 calibrations and fast deliveries!
Conducted RF Probes, Clamps, & Devices
Item | Capabilities |
EM101 | EM Clamp .15 to 1,000 MHz, 100 W max, incl: FTC 101 decoupling clamp |
F120-9A | Clamp-on injection probe, 10kHz to 230MHz, 125 CW, calibration fixture |
CDN-AF4 | 100v, 1.5amp. coupler for 4 unshielded lines |
CDN-S15 | 100V, 1.5 amp. CDN for 15 shielded lines |
CDN-M2/3/16A | 240VAC, 16A coupler for 2 or 3 power lines |
CDN-M5/32A |
400VAC, 32A coupler for 5 power lines |
Additional Available through Manufacturer |
|
MD 4070A | Monitoring Device 4kHz - 400MHz |
SW 4070, SW 4070-MAN | RF Switch NSG 4070 up to 2GHz |
ATN Series | 6 dB Attenuators |
Other Conducted RF Immunity Accessories
Item | Capabilities |
ATT6/75 | 6dB attenuator, 75W |
ATT20/0.5 | 20db attenuator, 0.5W for the calibration of CDNs and clamps (included with the PM 1000) |
CDNs | As per IEC 61000-4-6 (refer to separate list) |
Clamps | 50ohm termination resistor for CDNs and clamps |
T-50 | 240VAC, 16A coupler for 2 or 3 power lines |
R-100x |
150ohm-to-50ohm matching impedance for calibration |
Cal adapters |
For all types of CDNs and clamps |
icd.control | Extensive and most versatile remote control and reporting software. A standard library helps to configure the test setup. Multiple interruption functions automated by IEEE instruments or manually. Easy to use or expandable to complex test routines on the base of vector definitions. |
SW 4070, SW 4070-MAN | RF Switch NSG 4070 up to 2GHz |
ATN Series | 6 dB Attenuators |