LAS 6100 is a loop antenna set designed to generate and verify magnetic fields in close proximity to test the immunity of electrical equipment exposed to RF fields as required by the IEC 61000-4-39 standard.
The LAS 6100 antenna set consists of a radiating loop antenna RLA 6100-3, field sensor loop FSL 6040-1 and guarantees the required distance of 50 mm needed to comply with the standard requirements.
FSL 6040-1 field loop sensor with a diameter of 40 mm, offers electrostatic shielding. The correction factor is also provided with the FSL 6040-1.
The LAS 6100 set includes the ANP 4039 adaptor to allow more field strength at 13.56 MHz, as required by IEC 60601-1-2.
Data Sheet User ManualSpecifications
RLA 6100-3 Radiating Loop Antenna |
|
Frequency range |
150 kHz to 30 MHz |
Test distance |
50 mm |
Required Power |
see graph for typical drive power |
Loop diameter |
100 mm |
No. of turns |
120 mm |
No. of turns |
3 |
Wire diameter |
approx. 1 mm |
Connector type |
BNC, 50 Ω |
Max. input power |
100 W |
Dimensions (W x H x D): |
110 mm x 120 mm x 78 mm |
FSL 6040-1 Field Sensor Loop |
|
Frequency range |
150 kHz to 30 MHz |
Loop diameter |
40 mm |
No. of turns |
51 |
Wire diameter |
approx. 0.5 mm |
Shielding |
electrostatic |
Connector type |
BNC, 50 Ω |
Dimensions (W x H x D): |
50 mm x 120 mm x 25 mm |
ANP 4039 Matching Network 13.56 MH |
|
Applicable frequency |
13.56 MHz |
RF input connector type |
N, 50 Ω |
RF output connector type |
22SHV-50-0-2 (cable LE 271 required) |
Max. input power |
20 W |
Dimensions (W x H x D): |
90 mm x 63 mm x 110 mm |
Weight: |
approx. 0.5 kg |
Environment specifications |
|
Classification |
Indoor use only |
Operation temperature |
0 °C to +40 °C |
Storage and transport temperature |
-10 °C to +60 °C |
Relative humidity |
up to 90 % (no moisture condensation) |