EM Test UCS 500N7

Rent 7 kV Transient Test System
In Stock - Order by 3 p.m. ships today
  • 7 kV Surge, 5.5 kV EFT, & 6 kV 100 kHz Ringwave
  • Built-in CDN Max 300 V/16 A AC or 100 V/10 A DC
  • Includes Laptop with iec.control Software
  • Test Routines to Many Immunity Requirements
  • Current ISO 17025 Accredited Calibration
Order Type

Included Accessories

  • Immunity Generator
  • Power Cord
  • EUT In Supply Cord
  • ESA Adapte
  • Sync cord black
  • USB Cable
  • Software with Manual on USB Stick
  • Printed Manual
  • Calibration Documentation
  • Black U shaped jumper (attached in back)
  • Pelican Case 
Can't find something? Contact Us
Got Questions? Speak to an Expert +1 630-887-0329
Two FREE days transit with every rental

Description

The UCS 500N7 ultra-compact simulator is the most versatile tester to cover transient and power fail requirements according to international standards (basic and generic standards) and product/ product family standards with voltage capability of up to 7.0 kV. Apart from the IEC/EN 61000-4-5 standard for surge testing it also complies to ANSI/IEEE C62.41 for surge and ringwave testing.

 The UCS 500N7 not only represents the most economic solution for full-compliant immunity tests and CE Marking but goes far beyond. Having a built-in CDN for single phase DUTs it can be extended for testing three-phase DUTs by means of an automatically controlled external coupling network up to 100A. EM TEST supplies a large range of accessories for various applications.

Benefits

The UCS 500N7 includes everything necessary to conduct fully compliant tests at levels that go far beyond common test requirements.

Surge can be test up to 7.0kV either according to IEC/EN 61000-4-5 or ANSI/IEEE C62.41. The UCS 500N7 offers an integrated Ringwave module (optional) as per ANSI/IEEE C62.41 and IEC/EN 61000-4-12 for ringwave tests on mains supply lines. 

The UCS 500N7 can be operated manually from the front panel or by remote via the built-in USB or GPIB interface. Fail inputs allow to control an ongoing test sequence based on the status of the DUT. Monitoring outputs (BNC) are offered for easy signal measurement and verification. Safety features such as interlock and warning lamp control are available.

Pre-programmed Standard Test routines allow highest user convenience. Still the UCS 500N7 offers the Quick Start test routine where parameters can be changed on-line during the test to evaluate the susceptibility level of an individual DUT, a most appreciated benefit for tests at development stage.

Operation

Front panel menu and function keys enable the user to program his test routines quickly and accurately. The cursor allows fast control of all test parameters of the programmed routine, thus test procedures are simplified and confidence is generated that every step is carried out correctly.

Software

Outstanding user convenience, clearly structured windows and operation features and the EM TEST standards library along with the flexibility to generate user specific test sequences very easily are the main features of iec.control software.

The software is automatically configured according to the connected EM TEST generators. Extensive reporting capabilities help the user to create test reports that meet international requirements.

 iec.control is supported by Windows XP, Windows Vista, Windows 7 and Windows 8. Remote control is achieved either via USB or GPIB. iec.control supports a wide range of GPIB cards of National Instruments.

Specifications:
Burst Module
As per EN/IEC 61000-4-4 and EN 61000-6-1, -6-2
Test voltage
200V - 5,500 V +/- 10%;
100V - 2,750 V +/- 10% into 50 ohm
Pulse shape
5/50 ns into 50 ohm and 1,000 ohm
Rise time tr
5 ns +/- 30% into 50 ohm; 5 ns +/- 30% into 1,000 ohm
Pulse width td
50 ns +/- 30% into 50 ohm;
50 ns -15/+100 ns into 1,000 ohm
Source impedance
50 ohm
Polarity
Positive, negative
Trigger Circuit
Trigger of bursts
Automatic, manual, external
Synchronization
0 - 360 Degrees, resolution 1 degree (16 - 500 Hz)
Burst duration
td = 0.10 ms - 999 ms
Repetition rate
tr = 10 ms - 9,999 ms
Spike frequency
f = 0.1 kHz - 1,000 kHz
Test duration
T = 0:01 min - 99:59 min
T > 99:59 min --> endless
Outputs
Direct
Via 50 ohm coaxial connector
Coupling mode
L, N, PE; all combinations
DUT supply
AC: 300 V/16 A; 50/60 Hz
DC: 300 V/16 A
CRO trigger
5 V trigger signal for oscilloscope
Test Routines
Quick Start
On-line adjustable parameters, easy-to-use
Standard Test routines
As per IEC 61000-4-4,
Levels 1 - 4 As per EN 61000-6-1, -6-2
Manual Standard Test routine
User Test routines
Synchronous burst release, Random burst release, Change voltage after T, Frequency sweep within one burst,
Frequency sweep with constant number of pulses, Frequency sweep with constant burst duration, Change polarity after T
Surge Module
As per EN/IEC 61000-4-5 Ed 3.0 and EN 61000-6-1, -6-2
Voltage (o.c.)
250 V - 7,000 V +/- 10%
Pulse front time
1.2 us +/- 30%
Pulse time to half value
50 us +/- 20%
Current (s.c.)
Max. 3,500 A +/- 10%
Pulse front time
8 us +/- 20%
Pulse time to half value
20 us +/- 20%
Polarity
Positive, negative, alternating
Counter
1 - 30,000 or endless, selectable
Trigger
Release of pulses
Automatic, manual, external
Synchronization
0 - 360 degrees, resolution 1 degree
Repetition rate
Max. 0.5 Hz (2 s - 999 s)
Outputs
Direct
Via HV connectors for external coupling networks (Zi = 2 ohm with optional adapter IMN 2)
Coupling modes
As per IEC 61000-4-5: Line to line with 2 ohm, Line(s) to ground with 12 ohm
As per ANSI/IEEE C62.41: Line to line with 2ohm, Line(s) to ground with 2ohm
DUT supply
AC: 300 V/16 A; 50/60 Hz
DC: 300 V/16 A
CRO trigger
5 V trigger signal for oscilloscope
Measurements
CRO U-monitor
10 Vp at 7,000 V
CRO I-monitor
10 Vp at 3,500 A
Peak voltage
7,000 V in the LCD display
Peak current
3,500 A in the LCD display
Overcurrent protection
Breaks the Sureg test when the surge current is over the limit, Limitter for differential mode, Limitter for common mode
Test Routines
Quick Start
One-line adjustable parameters, easy-to-use
Standard Test routines
As per IEC 61000-4-5, Levels 1 - 4
As per EN 61000-6-1, -6-2
Manual Standard Test routine
User Test routines
Change polarity after n pulses
Change coupling after n pulses
Change voltage after n pulses
Change phase angle after n pulses
Pulsed Magnetic Field
As per IEC 61000-4-9
Test levels 100, 300 and 1,000 A/m
Test level steplessly adjustable under Quick Start
Ringwave Module
As per ANSI/IEEE C62.41 and EN/IEC 61000-4-12
Test voltage
250 V - 6,000 V +/- 10%
Voltage
Wave shape (open circuit)
Rise time (first peak)
0.5 us +/- 30%
Oscillation frequency
100 kHz +/- 10%
Decaying
Peak 2 to peak 1 = 40 - 110%
Peak 3 to peak 2 = 40 - 80%
Peak 4 to peak 3 = 40 - 80%
Current
Wave shape (short circuit)
Rise time
<= 1.0 us
Oscillation frequency
100 kHz +/- 10%
Source impedance
12 ohm and 30 ohm
Peak current
As per selected source impedance
Polarity
Positive, negative
Trigger
Release of pulses
Automatic, manual, external
Synchronization
0 - 360 degrees, resolution 1 degree
Repetition rate
max. 1 Hz (1 s - 999 s)
Outputs
Direct
Via HV-safety lab connectors
Coupling mode
L, N, PE; line to line and line(s) to ground
DUT supply
AC: 300 V/16 A; 50/60 Hz
DC: 300 V/16 A
CRO trigger
5 V trigger signal for oscilloscope
Quick Start
On-line adjustable parameters, easy-to-use