Teseq NSG 4070C2-LFCP - Rent Close Proximity Magnetic Field System

Teseq Magnetic Field Test System:

  • 100 Watts 9 kHz to 50 MHz Power Amplifier
  • Testing to IEC 61000-4-39 & IEC 60601-1-2
  • Rental Antenna Test Sets Available
  • Includes Laptop with icd.control Software
  • Current ISO 17025 Calibration Including Data

Antenna Kits


Get weekly, two week, and monthly rentals on all products

Got Questions about the equipment? Get technical support

Two FREE days transit time with every rental

Request A Quote

Related Test Equipment

Description

Overview

The NSG 4070C-LFCP is a universal device for standard-compliant and development-accompanying EMC immunity tests in the application area of magnetic field testing in the close proximity according to IEC / EN 61000-4-39 and IEC / EN 60601-1-2. The NSG 4070C-LFCP integrates signal generator, directional coupler, power amplifier, power meter and EUT monitoring interfaces. The NSG 4070C-LFCP can be operated quickly, conveniently and easily via the front panel as a free-standing device. Test and measurement data can be conveniently transferred for documentation purposes via a USB stick. The NSG 4070C-LFCP can be remote controlled via LAN, electrical or optical RS232 as well as USB.

The wide frequency range of the signal generator and the power meter offer the possibility to connect external amplifiers and directional couplers. In combination with a remote control software, external power amplifiers and directional couplers, further applications can be covered, such as testing of conducted disturbances induced by high frequency fields according to IEC / EN 61000-4-6, automotive BCI tests e.g. according to ISO 11452-4 as well as high frequency electromagnetic fields according to IEC / EN 61000-4-3 or IEC / EN 61000-4-20. For EUT monitoring, the NSG 4070C-LFCP offers a variety of interfaces for flexibility in laboratory use.

In order to start with predefined parameter settings the optional test software icd.control is recommended. The software offers a large standard database and predefined drives for using external measuring devices. More complex systems including radiated tests can be controlled by using the software solution CIS (Compliance Immunity Software).t also allows for all coupling methods to be simply and easily connected and calibrated.

Data SheetTesting BrochureUser Manual

Specifications

Generator


RF

 

 Frequency Range

4 kHz to 1 GHz

 Resolution

1 Hz

 Reference Frequency

10 MHz

 Aging

25 ppm

RF Level


 Level range:

-60 dBm to +10 dBm

 Resolution:

0.1 dB

 Settling time:

10 ms

Amplitude modulation

 

 Modulation depth:

0 to 100%

 Modulation frequency range:

1 Hz to 50 kHz

 Frequency resolution:

1 Hz

External modulation


 Delay time:

< 1 µs /180°

 Period:

min. 20 µs

 Pulse width:

 min. 10 µs

Power meter


 Frequency range:

4 kHz to 1 GHz

 Linear measurement range

  channel 1:

-35 dBm to +27 dBm

  channel 2,3:

-45 dBm to +20 dBm

Max. input/no damage


  channel 1-3:

+28 dBm

Noise level:

>5 dB below the measurement rang

Input return loss:

>20 dB (below 500 MHz), >17 dB (500 MHz to 1 GHz)

Connector:

BNC socket, 50 Ω

Accuracy 10 to 30°C:

<0.5 dB, typ. <0.3 dB

Power Amplifier


Nominal output power:

100 W

Frequency range:

9 kHz - 50 MHz

Type:

single band, class A

Input/output impedance (nominal):

50 Ω

Input return loss (minimum):

10 dB

Output return loss (nominal):

min. 9.5 dB

Output return loss without damage:

0 dB

Gain (minimum):

50 dB