Description
Combination Wave Surge System Features:
- IEC 61000-4-5 & ANSI C37.90 Solution
- ISO 17025 Calibration Included
- 4.4kV Generator & Built-in Pulse Monitor
- Includes 3-Phase CDN & E103 Controller
- Free Technical Support & 2 Days Free Transit
Overview
The ECAT E103, E501, and E4554 by Thermo Fisher Scientific / Keytek is designed to test to many EMC immunity standards requiring 6kV surge/combination wave with 3 phase power coupler/decoupler 480v, 32a EUT. This ECAT System comes with an A2LA accredited calibration and free technical support included with rental. This rental is a combination of the ECAT E103 control unit and E501, and E4554 modules.
This Thermo Fisher Scientific / Keytek EMC immunity generator provides single port testing and allows changing tests modes without switching off power to the equipment under test (EUT). This helps testing with to be conducted efficiently without re-initializing the EUT between tests. The ECAT E501 can be rented individually or in combination with the ECAT E103.
This ECAT system is designed to test EN/IEC 61000-4-5 ed.3, ANSI/IEEE C62.41, and UL 1449 and many other immunity standards that have a high voltage surge component. The modular system of this Thermo Fisher / Keytek product provides the ability to be upgraded providing the ability to upgrade test capabilities to meet the newest standard. The E103 and E501, E4554 combination can be controlled via front panel or through computer using Keytek software.
Specifications
Controller-E103 |
Controller for the ECAT system with front panel and software operation. |
Combination Wave Surge-E501B: |
|
Standards |
IEC 61000-4-5 ed. 3, EN 61000-4-5, ANSI/IEEE C62.41.2 Cat. B, UL 1449 at 3kA |
Open Circuit Voltage Waveform |
1.2µs, Front Time ± 30%, 50µs Duration ±20% |
Short Circuit Current Waveform |
8µs Front Time ±20%, 20µs Duration ±20% |
Peak Open Circuit Voltage |
250V–6kV ± 10% @2 Ω, 250-6kV ±10% @12 Ω |
Peak Short Circuit Current |
125A–3kA ±10%@2 Ω, 20.8-500A ±10%@12Ω |
Pulse Repetition Rate |
2 pulses/min. @6kV |
100 kHz Ringwave-E503: |
|
Standards |
ANSI/IEEE C62.41.2 Cat. A& B, UL 864, EN IEC 61000-4-2, ANSI C63.16 |
Open Circuit Voltage Waveform |
100kHz Damped Cosine, Rise Time=.5µs±30%, Damping= 40%/peak |
Peak Open Circuit Voltage |
200V–6.6kV ± 10% @12 Ω or @ 30 Ω, Front Time0.5µs ±30% |
Peak Short Circuit Current |
200A ±10%@30 Ω, 500A±10%@12 Ω |
Pulse Repetition Rate |
6 pulses/min. |
Line Sink Accuracy |
± 15° with optional coupler/decoupler |
Integrated CDN-E4554: |
|
AC Voltage and Current |
50-480 VAC, 32A, 5 lines/3 phases |
DC Voltage and Current |
N/A |
Testimonials
I would recommend Transient Specialists. Transient Specialists always provides quality products within the established timeframes. The equipment always meets the specifications and properly shipped within the time frames within the time frames promised.
Scott D., TUV SUD Canada
Transient Specialists has been outstanding at working with us for our specialized EMC equipment rental needs. Fast shipments and excellent accommodation for some customized setups. I recommend them highly and they are a pleasure to partner with!
John H., Gates Air, Inc.
Laura, I want to thank you and Goran for your excellent customer service and turn around time on the rentals and the calibration documents.
Adam P., Clark Testing
Working for a testing company I deal with renting equipment all the time. Transient Specialist is the first company I call on my list of rental houses because of their great personnel, prompt delivery of the equipment, knowledgeable and accommodating to the customers needs.
Richard T., National Technical Systems